Performance of a Small-Sized, Critical Flowmeter Assembly
Measurement of low mass-flow rates was readily accomplished by development of two similar flowmetering assemblies. Each assembly consisted of five 0.016 to 0.070-in. diameter critical-flow nozzles and individual control valves connected between common inlet and exit manifolds. Mass-flow rates throug...
Saved in:
Published in: | Journal of basic engineering 1965-12, Vol.87 (4), p.1025-1027 |
---|---|
Main Author: | |
Format: | Article |
Language: | English |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Measurement of low mass-flow rates was readily accomplished by development of two similar flowmetering assemblies. Each assembly consisted of five 0.016 to 0.070-in. diameter critical-flow nozzles and individual control valves connected between common inlet and exit manifolds. Mass-flow rates through each nozzle were determined over a range of upstream pressure and temperature. The probable deviations in the measured data are approximately ≠0.83 percent. Discharge coefficients were calculated from the calibration results using the well-known one-dimensional critical-flow equation. An average error of ≠2.5 percent in the discharge coefficient is possible because of the difficulty in determining true throat areas of these small-sized nozzles. |
---|---|
ISSN: | 0098-2202 0021-9223 1528-901X |
DOI: | 10.1115/1.3650794 |