Electrosurgery and temperature increase in tissue with a passive metal implant

Importance: During monopolar electrosurgery in patients, current paths can be influenced by metal implants, which can cause unintentional tissue heating in proximity to implants. Guidelines concerning electrosurgery and active implants such as pacemakers or implantable cardioverter defibrillators ha...

Full description

Saved in:
Bibliographic Details
Published in:Frontiers in surgery 2019
Main Authors: Martinsen, Tormod, Pettersen, Fred-Johan, Kalvøy, Håvard, Tronstad, Christian, Kvarstein, Gunnvald, Bakken, André, Høgetveit, Jan Olav, Martinsen, Ørjan Grøttem, Grimnes, Sverre, Frich, Lars Inge
Format: Article
Language:Norwegian
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by
cites
container_end_page
container_issue
container_start_page
container_title Frontiers in surgery
container_volume
creator Martinsen, Tormod
Pettersen, Fred-Johan
Kalvøy, Håvard
Tronstad, Christian
Kvarstein, Gunnvald
Bakken, André
Høgetveit, Jan Olav
Martinsen, Ørjan Grøttem
Grimnes, Sverre
Frich, Lars Inge
description Importance: During monopolar electrosurgery in patients, current paths can be influenced by metal implants, which can cause unintentional tissue heating in proximity to implants. Guidelines concerning electrosurgery and active implants such as pacemakers or implantable cardioverter defibrillators have been published, but most describe interference between electrosurgery and the active implant rather than the risk of unintended tissue heating. Tissue heating in proximity to implants during electrosurgery may cause an increased risk of patient injury. Objective: To determine the temperature of tissue close to metal implants during electrosurgery in an in-vitro model. Design, Setting, and Participants: Thirty tissue samples (15 with a metal implant placed in center, 15 controls without implant) were placed in an in vitro measurement chamber. Electrosurgery was applied at 5–60 W with the active electrode at three defined distances from the implant while temperatures at four defined distances from the implant were measured using fiber-optic sensors. Main Outcomes and Measures: Tissue temperature increase at the four tissue sites was determined for all power levels and each of the electrode-to-implant distances. Based on a linear mixed effects model analysis, the primary outcomes were the difference in temperature increase between implant and control tissue, and the estimated temperature increase per watt per minute. Results: Tissues with an implant had higher temperature increases than controls at all power levels after 1 min of applied electrosurgery (mean difference of 0.16°C at 5 W, 0.50°C at 15 W, 1.11°C at 30 W, and 2.22°C at 60 W, all with p < 0.001). Temperature increase close to the implant was estimated to be 0.088°C/W/min (95% CI: 0.078–0.099°C/W/min; p < 0.001). Temperature could increase to above 43°C after 1 min of 60 W. Active electrode position had no significant effect on temperature increases for tissues with implant (p = 0.6). Conclusions and Relevance: The temperature of tissue close to a metal implant increases with passing electrosurgery current. There is a significant risk of high tissue temperature when long activation times or high power levels are used.
format article
fullrecord <record><control><sourceid>cristin</sourceid><recordid>TN_cdi_cristin_nora_10852_75842</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>10852_75842</sourcerecordid><originalsourceid>FETCH-cristin_nora_10852_758423</originalsourceid><addsrcrecordid>eNqFyj0KwkAQQOFFFBTNGZwLBDabX2uJWFlZ2IUhjrqy2YSZjeLtRbCws3pf8SZqYcymiKsyP01_PFeRyF1rnaRZUphsoQ61ozZwLyNfiV-A_gyBuoEYw8gE1rdMKB9AsCIjwdOGGyAMKGIfBB0FdGC7waEPKzW7oBOKvl2q9a4-bvdxy1aC9Y3vGZtEV7lpyrzKTPr_eAPdGT1n</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>Electrosurgery and temperature increase in tissue with a passive metal implant</title><source>NORA - Norwegian Open Research Archives</source><source>PubMed Central</source><creator>Martinsen, Tormod ; Pettersen, Fred-Johan ; Kalvøy, Håvard ; Tronstad, Christian ; Kvarstein, Gunnvald ; Bakken, André ; Høgetveit, Jan Olav ; Martinsen, Ørjan Grøttem ; Grimnes, Sverre ; Frich, Lars Inge</creator><creatorcontrib>Martinsen, Tormod ; Pettersen, Fred-Johan ; Kalvøy, Håvard ; Tronstad, Christian ; Kvarstein, Gunnvald ; Bakken, André ; Høgetveit, Jan Olav ; Martinsen, Ørjan Grøttem ; Grimnes, Sverre ; Frich, Lars Inge</creatorcontrib><description>Importance: During monopolar electrosurgery in patients, current paths can be influenced by metal implants, which can cause unintentional tissue heating in proximity to implants. Guidelines concerning electrosurgery and active implants such as pacemakers or implantable cardioverter defibrillators have been published, but most describe interference between electrosurgery and the active implant rather than the risk of unintended tissue heating. Tissue heating in proximity to implants during electrosurgery may cause an increased risk of patient injury. Objective: To determine the temperature of tissue close to metal implants during electrosurgery in an in-vitro model. Design, Setting, and Participants: Thirty tissue samples (15 with a metal implant placed in center, 15 controls without implant) were placed in an in vitro measurement chamber. Electrosurgery was applied at 5–60 W with the active electrode at three defined distances from the implant while temperatures at four defined distances from the implant were measured using fiber-optic sensors. Main Outcomes and Measures: Tissue temperature increase at the four tissue sites was determined for all power levels and each of the electrode-to-implant distances. Based on a linear mixed effects model analysis, the primary outcomes were the difference in temperature increase between implant and control tissue, and the estimated temperature increase per watt per minute. Results: Tissues with an implant had higher temperature increases than controls at all power levels after 1 min of applied electrosurgery (mean difference of 0.16°C at 5 W, 0.50°C at 15 W, 1.11°C at 30 W, and 2.22°C at 60 W, all with p &lt; 0.001). Temperature increase close to the implant was estimated to be 0.088°C/W/min (95% CI: 0.078–0.099°C/W/min; p &lt; 0.001). Temperature could increase to above 43°C after 1 min of 60 W. Active electrode position had no significant effect on temperature increases for tissues with implant (p = 0.6). Conclusions and Relevance: The temperature of tissue close to a metal implant increases with passing electrosurgery current. There is a significant risk of high tissue temperature when long activation times or high power levels are used.</description><identifier>ISSN: 2296-875X</identifier><identifier>EISSN: 2296-875X</identifier><language>nor</language><publisher>Frontiers Media</publisher><ispartof>Frontiers in surgery, 2019</ispartof><rights>info:eu-repo/semantics/openAccess</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>230,314,780,784,885,4024,26567</link.rule.ids></links><search><creatorcontrib>Martinsen, Tormod</creatorcontrib><creatorcontrib>Pettersen, Fred-Johan</creatorcontrib><creatorcontrib>Kalvøy, Håvard</creatorcontrib><creatorcontrib>Tronstad, Christian</creatorcontrib><creatorcontrib>Kvarstein, Gunnvald</creatorcontrib><creatorcontrib>Bakken, André</creatorcontrib><creatorcontrib>Høgetveit, Jan Olav</creatorcontrib><creatorcontrib>Martinsen, Ørjan Grøttem</creatorcontrib><creatorcontrib>Grimnes, Sverre</creatorcontrib><creatorcontrib>Frich, Lars Inge</creatorcontrib><title>Electrosurgery and temperature increase in tissue with a passive metal implant</title><title>Frontiers in surgery</title><description>Importance: During monopolar electrosurgery in patients, current paths can be influenced by metal implants, which can cause unintentional tissue heating in proximity to implants. Guidelines concerning electrosurgery and active implants such as pacemakers or implantable cardioverter defibrillators have been published, but most describe interference between electrosurgery and the active implant rather than the risk of unintended tissue heating. Tissue heating in proximity to implants during electrosurgery may cause an increased risk of patient injury. Objective: To determine the temperature of tissue close to metal implants during electrosurgery in an in-vitro model. Design, Setting, and Participants: Thirty tissue samples (15 with a metal implant placed in center, 15 controls without implant) were placed in an in vitro measurement chamber. Electrosurgery was applied at 5–60 W with the active electrode at three defined distances from the implant while temperatures at four defined distances from the implant were measured using fiber-optic sensors. Main Outcomes and Measures: Tissue temperature increase at the four tissue sites was determined for all power levels and each of the electrode-to-implant distances. Based on a linear mixed effects model analysis, the primary outcomes were the difference in temperature increase between implant and control tissue, and the estimated temperature increase per watt per minute. Results: Tissues with an implant had higher temperature increases than controls at all power levels after 1 min of applied electrosurgery (mean difference of 0.16°C at 5 W, 0.50°C at 15 W, 1.11°C at 30 W, and 2.22°C at 60 W, all with p &lt; 0.001). Temperature increase close to the implant was estimated to be 0.088°C/W/min (95% CI: 0.078–0.099°C/W/min; p &lt; 0.001). Temperature could increase to above 43°C after 1 min of 60 W. Active electrode position had no significant effect on temperature increases for tissues with implant (p = 0.6). Conclusions and Relevance: The temperature of tissue close to a metal implant increases with passing electrosurgery current. There is a significant risk of high tissue temperature when long activation times or high power levels are used.</description><issn>2296-875X</issn><issn>2296-875X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2019</creationdate><recordtype>article</recordtype><sourceid>3HK</sourceid><recordid>eNqFyj0KwkAQQOFFFBTNGZwLBDabX2uJWFlZ2IUhjrqy2YSZjeLtRbCws3pf8SZqYcymiKsyP01_PFeRyF1rnaRZUphsoQ61ozZwLyNfiV-A_gyBuoEYw8gE1rdMKB9AsCIjwdOGGyAMKGIfBB0FdGC7waEPKzW7oBOKvl2q9a4-bvdxy1aC9Y3vGZtEV7lpyrzKTPr_eAPdGT1n</recordid><startdate>2019</startdate><enddate>2019</enddate><creator>Martinsen, Tormod</creator><creator>Pettersen, Fred-Johan</creator><creator>Kalvøy, Håvard</creator><creator>Tronstad, Christian</creator><creator>Kvarstein, Gunnvald</creator><creator>Bakken, André</creator><creator>Høgetveit, Jan Olav</creator><creator>Martinsen, Ørjan Grøttem</creator><creator>Grimnes, Sverre</creator><creator>Frich, Lars Inge</creator><general>Frontiers Media</general><scope>3HK</scope></search><sort><creationdate>2019</creationdate><title>Electrosurgery and temperature increase in tissue with a passive metal implant</title><author>Martinsen, Tormod ; Pettersen, Fred-Johan ; Kalvøy, Håvard ; Tronstad, Christian ; Kvarstein, Gunnvald ; Bakken, André ; Høgetveit, Jan Olav ; Martinsen, Ørjan Grøttem ; Grimnes, Sverre ; Frich, Lars Inge</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-cristin_nora_10852_758423</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>nor</language><creationdate>2019</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Martinsen, Tormod</creatorcontrib><creatorcontrib>Pettersen, Fred-Johan</creatorcontrib><creatorcontrib>Kalvøy, Håvard</creatorcontrib><creatorcontrib>Tronstad, Christian</creatorcontrib><creatorcontrib>Kvarstein, Gunnvald</creatorcontrib><creatorcontrib>Bakken, André</creatorcontrib><creatorcontrib>Høgetveit, Jan Olav</creatorcontrib><creatorcontrib>Martinsen, Ørjan Grøttem</creatorcontrib><creatorcontrib>Grimnes, Sverre</creatorcontrib><creatorcontrib>Frich, Lars Inge</creatorcontrib><collection>NORA - Norwegian Open Research Archives</collection><jtitle>Frontiers in surgery</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Martinsen, Tormod</au><au>Pettersen, Fred-Johan</au><au>Kalvøy, Håvard</au><au>Tronstad, Christian</au><au>Kvarstein, Gunnvald</au><au>Bakken, André</au><au>Høgetveit, Jan Olav</au><au>Martinsen, Ørjan Grøttem</au><au>Grimnes, Sverre</au><au>Frich, Lars Inge</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Electrosurgery and temperature increase in tissue with a passive metal implant</atitle><jtitle>Frontiers in surgery</jtitle><date>2019</date><risdate>2019</risdate><issn>2296-875X</issn><eissn>2296-875X</eissn><abstract>Importance: During monopolar electrosurgery in patients, current paths can be influenced by metal implants, which can cause unintentional tissue heating in proximity to implants. Guidelines concerning electrosurgery and active implants such as pacemakers or implantable cardioverter defibrillators have been published, but most describe interference between electrosurgery and the active implant rather than the risk of unintended tissue heating. Tissue heating in proximity to implants during electrosurgery may cause an increased risk of patient injury. Objective: To determine the temperature of tissue close to metal implants during electrosurgery in an in-vitro model. Design, Setting, and Participants: Thirty tissue samples (15 with a metal implant placed in center, 15 controls without implant) were placed in an in vitro measurement chamber. Electrosurgery was applied at 5–60 W with the active electrode at three defined distances from the implant while temperatures at four defined distances from the implant were measured using fiber-optic sensors. Main Outcomes and Measures: Tissue temperature increase at the four tissue sites was determined for all power levels and each of the electrode-to-implant distances. Based on a linear mixed effects model analysis, the primary outcomes were the difference in temperature increase between implant and control tissue, and the estimated temperature increase per watt per minute. Results: Tissues with an implant had higher temperature increases than controls at all power levels after 1 min of applied electrosurgery (mean difference of 0.16°C at 5 W, 0.50°C at 15 W, 1.11°C at 30 W, and 2.22°C at 60 W, all with p &lt; 0.001). Temperature increase close to the implant was estimated to be 0.088°C/W/min (95% CI: 0.078–0.099°C/W/min; p &lt; 0.001). Temperature could increase to above 43°C after 1 min of 60 W. Active electrode position had no significant effect on temperature increases for tissues with implant (p = 0.6). Conclusions and Relevance: The temperature of tissue close to a metal implant increases with passing electrosurgery current. There is a significant risk of high tissue temperature when long activation times or high power levels are used.</abstract><pub>Frontiers Media</pub><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 2296-875X
ispartof Frontiers in surgery, 2019
issn 2296-875X
2296-875X
language nor
recordid cdi_cristin_nora_10852_75842
source NORA - Norwegian Open Research Archives; PubMed Central
title Electrosurgery and temperature increase in tissue with a passive metal implant
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-28T14%3A35%3A49IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-cristin&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Electrosurgery%20and%20temperature%20increase%20in%20tissue%20with%20a%20passive%20metal%20implant&rft.jtitle=Frontiers%20in%20surgery&rft.au=Martinsen,%20Tormod&rft.date=2019&rft.issn=2296-875X&rft.eissn=2296-875X&rft_id=info:doi/&rft_dat=%3Ccristin%3E10852_75842%3C/cristin%3E%3Cgrp_id%3Ecdi_FETCH-cristin_nora_10852_758423%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true