Loading…

A hybrid multi-particle approach to range assessment-based treatment verification in particle therapy

Particle therapy (PT) used for cancer treatment can spare healthy tissue and reduce treatment toxicity. However, full exploitation of the dosimetric advantages of PT is not yet possible due to range uncertainties, warranting development of range-monitoring techniques. This study proposes a novel ran...

Full description

Saved in:
Bibliographic Details
Main Authors: Meric, Ilker, Alagoz, Enver, Hysing, Liv Bolstad, Kögler, Toni, Lathouwers, Danny, Lionheart, William R. B, Mattingly, John, Obhodas, Jasmina, Pausch, Guntram, Pettersen, Helge Egil Seime, Ratliff, Hunter Nathaniel, Rovituso, Marta, Schellhammer, Sonja M, Setterdahl, Lena Marie, Skjerdal, Kyrre, Sterpin, Edmond, Sudac, Davorin, Turko, Joseph A, Ytre-Hauge, Kristian Smeland
Format: Article
Language:English
Online Access:Request full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Particle therapy (PT) used for cancer treatment can spare healthy tissue and reduce treatment toxicity. However, full exploitation of the dosimetric advantages of PT is not yet possible due to range uncertainties, warranting development of range-monitoring techniques. This study proposes a novel range-monitoring technique introducing the yet unexplored concept of simultaneous detection and imaging of fast neutrons and prompt-gamma rays produced in beam-tissue interactions. A quasimonolithic organic detector array is proposed, and its feasibility for detecting range shifts in the context of proton therapy is explored through Monte Carlo simulations of realistic patient models and detector resolution efects. The results indicate that range shifts of 1 mm can be detected at relatively low proton intensities (22.30(13) × 107 protons/spot) when spatial information obtained through imaging of both particle species are used simultaneously. This study lays the foundation for multiparticle detection and imaging systems in the context of range verifcation in PT