Loading…

An electrochemical biosensor for the determination of hormone Human Chorionic Gonadotropin (hCG) in human serum

This work describes the modification of a gold electrode to create an electrochemical biosensor capable of detecting human chorionic gonadotropin (hCG). The biosensor was obtained by modifying the gold electrode with cysteamine and oligopeptide (PPLRINRHILTR). The modification steps of the gold elec...

Full description

Saved in:
Bibliographic Details
Published in:Electroanalysis (New York, N.Y.) N.Y.), 2023-11, Vol.35 (11)
Main Authors: Koterwa, Adrian, Bojko, Magdalena, Ryl, Jacek, Łukaszuk, Krzysztof, Kozłowska, Kornelia, Sieklicki, Wiktor, Rodziewicz‐Motowidło, Sylwia, Niedziałkowski, Paweł
Format: Article
Language:English
Citations: Items that this one cites
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by
cites cdi_FETCH-LOGICAL-c194t-e94b52a9dee55a0fa3fbae7814f52607d8b84fd1a32f8e1038597a783e7522393
container_end_page
container_issue 11
container_start_page
container_title Electroanalysis (New York, N.Y.)
container_volume 35
creator Koterwa, Adrian
Bojko, Magdalena
Ryl, Jacek
Łukaszuk, Krzysztof
Kozłowska, Kornelia
Sieklicki, Wiktor
Rodziewicz‐Motowidło, Sylwia
Niedziałkowski, Paweł
description This work describes the modification of a gold electrode to create an electrochemical biosensor capable of detecting human chorionic gonadotropin (hCG). The biosensor was obtained by modifying the gold electrode with cysteamine and oligopeptide (PPLRINRHILTR). The modification steps of the gold electrode were confirmed by cyclic voltammetry (CV) and impedance electrochemical spectroscopy (EIS) measurements. The conducted EIS experiments in 0.01 M PBS, pH 7.4 confirm that the biosensor exhibits sensitivity towards hCG in a range of concentrations from 1×10 −12 to 1×10 −7  M (0.5 mIU/mL – 50 000 mIU/mL) to solutions with a detection limit of 1. 91×10 −14  M (0.0095 mIU/mL). The effectiveness of the investigated biosensor was also investigated in human serum. The EIS comparative investigations were performed in human serum containing a concentration of 1×10 −12  M (0.5 mIU/mL) hCG and in human serum where the hCG was added. The obtained results indicate that the investigated biosensor is selective for the presence of hCG hormone in the human serum.
doi_str_mv 10.1002/elan.202300095
format article
fullrecord <record><control><sourceid>crossref</sourceid><recordid>TN_cdi_crossref_primary_10_1002_elan_202300095</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>10_1002_elan_202300095</sourcerecordid><originalsourceid>FETCH-LOGICAL-c194t-e94b52a9dee55a0fa3fbae7814f52607d8b84fd1a32f8e1038597a783e7522393</originalsourceid><addsrcrecordid>eNo9kL1PwzAUxC0EEqWwMnuEIeH5q4nHKoIUqRILzJGTPCtGiV3Z6cB_TwKI4XT39E43_Ai5Z5AzAP6Eo_E5By4AQKsLsmGKs0wy0JdLBgkZCF1ck5uUPtfKTuoNCXtPccRujqEbcHKdGWnrQkKfQqR20Twg7XHGODlvZhc8DZYOIU7BIz2cJ-NptZzLw3W0Dt70YRk7OU8fhqp-pEsYfloJ43m6JVfWjAnv_nxLPl6e36tDdnyrX6v9MeuYlnOGWraKG90jKmXAGmFbg0XJpFV8B0VftqW0PTOC2xIZiFLpwhSlwEJxLrTYkvx3t4shpYi2OUU3mfjVMGhWXM2Kq_nHJb4BHn5fkQ</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>An electrochemical biosensor for the determination of hormone Human Chorionic Gonadotropin (hCG) in human serum</title><source>Wiley-Blackwell Read &amp; Publish Collection</source><creator>Koterwa, Adrian ; Bojko, Magdalena ; Ryl, Jacek ; Łukaszuk, Krzysztof ; Kozłowska, Kornelia ; Sieklicki, Wiktor ; Rodziewicz‐Motowidło, Sylwia ; Niedziałkowski, Paweł</creator><creatorcontrib>Koterwa, Adrian ; Bojko, Magdalena ; Ryl, Jacek ; Łukaszuk, Krzysztof ; Kozłowska, Kornelia ; Sieklicki, Wiktor ; Rodziewicz‐Motowidło, Sylwia ; Niedziałkowski, Paweł</creatorcontrib><description>This work describes the modification of a gold electrode to create an electrochemical biosensor capable of detecting human chorionic gonadotropin (hCG). The biosensor was obtained by modifying the gold electrode with cysteamine and oligopeptide (PPLRINRHILTR). The modification steps of the gold electrode were confirmed by cyclic voltammetry (CV) and impedance electrochemical spectroscopy (EIS) measurements. The conducted EIS experiments in 0.01 M PBS, pH 7.4 confirm that the biosensor exhibits sensitivity towards hCG in a range of concentrations from 1×10 −12 to 1×10 −7  M (0.5 mIU/mL – 50 000 mIU/mL) to solutions with a detection limit of 1. 91×10 −14  M (0.0095 mIU/mL). The effectiveness of the investigated biosensor was also investigated in human serum. The EIS comparative investigations were performed in human serum containing a concentration of 1×10 −12  M (0.5 mIU/mL) hCG and in human serum where the hCG was added. The obtained results indicate that the investigated biosensor is selective for the presence of hCG hormone in the human serum.</description><identifier>ISSN: 1040-0397</identifier><identifier>EISSN: 1521-4109</identifier><identifier>DOI: 10.1002/elan.202300095</identifier><language>eng</language><ispartof>Electroanalysis (New York, N.Y.), 2023-11, Vol.35 (11)</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><cites>FETCH-LOGICAL-c194t-e94b52a9dee55a0fa3fbae7814f52607d8b84fd1a32f8e1038597a783e7522393</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925</link.rule.ids></links><search><creatorcontrib>Koterwa, Adrian</creatorcontrib><creatorcontrib>Bojko, Magdalena</creatorcontrib><creatorcontrib>Ryl, Jacek</creatorcontrib><creatorcontrib>Łukaszuk, Krzysztof</creatorcontrib><creatorcontrib>Kozłowska, Kornelia</creatorcontrib><creatorcontrib>Sieklicki, Wiktor</creatorcontrib><creatorcontrib>Rodziewicz‐Motowidło, Sylwia</creatorcontrib><creatorcontrib>Niedziałkowski, Paweł</creatorcontrib><title>An electrochemical biosensor for the determination of hormone Human Chorionic Gonadotropin (hCG) in human serum</title><title>Electroanalysis (New York, N.Y.)</title><description>This work describes the modification of a gold electrode to create an electrochemical biosensor capable of detecting human chorionic gonadotropin (hCG). The biosensor was obtained by modifying the gold electrode with cysteamine and oligopeptide (PPLRINRHILTR). The modification steps of the gold electrode were confirmed by cyclic voltammetry (CV) and impedance electrochemical spectroscopy (EIS) measurements. The conducted EIS experiments in 0.01 M PBS, pH 7.4 confirm that the biosensor exhibits sensitivity towards hCG in a range of concentrations from 1×10 −12 to 1×10 −7  M (0.5 mIU/mL – 50 000 mIU/mL) to solutions with a detection limit of 1. 91×10 −14  M (0.0095 mIU/mL). The effectiveness of the investigated biosensor was also investigated in human serum. The EIS comparative investigations were performed in human serum containing a concentration of 1×10 −12  M (0.5 mIU/mL) hCG and in human serum where the hCG was added. The obtained results indicate that the investigated biosensor is selective for the presence of hCG hormone in the human serum.</description><issn>1040-0397</issn><issn>1521-4109</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2023</creationdate><recordtype>article</recordtype><recordid>eNo9kL1PwzAUxC0EEqWwMnuEIeH5q4nHKoIUqRILzJGTPCtGiV3Z6cB_TwKI4XT39E43_Ai5Z5AzAP6Eo_E5By4AQKsLsmGKs0wy0JdLBgkZCF1ck5uUPtfKTuoNCXtPccRujqEbcHKdGWnrQkKfQqR20Twg7XHGODlvZhc8DZYOIU7BIz2cJ-NptZzLw3W0Dt70YRk7OU8fhqp-pEsYfloJ43m6JVfWjAnv_nxLPl6e36tDdnyrX6v9MeuYlnOGWraKG90jKmXAGmFbg0XJpFV8B0VftqW0PTOC2xIZiFLpwhSlwEJxLrTYkvx3t4shpYi2OUU3mfjVMGhWXM2Kq_nHJb4BHn5fkQ</recordid><startdate>202311</startdate><enddate>202311</enddate><creator>Koterwa, Adrian</creator><creator>Bojko, Magdalena</creator><creator>Ryl, Jacek</creator><creator>Łukaszuk, Krzysztof</creator><creator>Kozłowska, Kornelia</creator><creator>Sieklicki, Wiktor</creator><creator>Rodziewicz‐Motowidło, Sylwia</creator><creator>Niedziałkowski, Paweł</creator><scope>AAYXX</scope><scope>CITATION</scope></search><sort><creationdate>202311</creationdate><title>An electrochemical biosensor for the determination of hormone Human Chorionic Gonadotropin (hCG) in human serum</title><author>Koterwa, Adrian ; Bojko, Magdalena ; Ryl, Jacek ; Łukaszuk, Krzysztof ; Kozłowska, Kornelia ; Sieklicki, Wiktor ; Rodziewicz‐Motowidło, Sylwia ; Niedziałkowski, Paweł</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c194t-e94b52a9dee55a0fa3fbae7814f52607d8b84fd1a32f8e1038597a783e7522393</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2023</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Koterwa, Adrian</creatorcontrib><creatorcontrib>Bojko, Magdalena</creatorcontrib><creatorcontrib>Ryl, Jacek</creatorcontrib><creatorcontrib>Łukaszuk, Krzysztof</creatorcontrib><creatorcontrib>Kozłowska, Kornelia</creatorcontrib><creatorcontrib>Sieklicki, Wiktor</creatorcontrib><creatorcontrib>Rodziewicz‐Motowidło, Sylwia</creatorcontrib><creatorcontrib>Niedziałkowski, Paweł</creatorcontrib><collection>CrossRef</collection><jtitle>Electroanalysis (New York, N.Y.)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Koterwa, Adrian</au><au>Bojko, Magdalena</au><au>Ryl, Jacek</au><au>Łukaszuk, Krzysztof</au><au>Kozłowska, Kornelia</au><au>Sieklicki, Wiktor</au><au>Rodziewicz‐Motowidło, Sylwia</au><au>Niedziałkowski, Paweł</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>An electrochemical biosensor for the determination of hormone Human Chorionic Gonadotropin (hCG) in human serum</atitle><jtitle>Electroanalysis (New York, N.Y.)</jtitle><date>2023-11</date><risdate>2023</risdate><volume>35</volume><issue>11</issue><issn>1040-0397</issn><eissn>1521-4109</eissn><abstract>This work describes the modification of a gold electrode to create an electrochemical biosensor capable of detecting human chorionic gonadotropin (hCG). The biosensor was obtained by modifying the gold electrode with cysteamine and oligopeptide (PPLRINRHILTR). The modification steps of the gold electrode were confirmed by cyclic voltammetry (CV) and impedance electrochemical spectroscopy (EIS) measurements. The conducted EIS experiments in 0.01 M PBS, pH 7.4 confirm that the biosensor exhibits sensitivity towards hCG in a range of concentrations from 1×10 −12 to 1×10 −7  M (0.5 mIU/mL – 50 000 mIU/mL) to solutions with a detection limit of 1. 91×10 −14  M (0.0095 mIU/mL). The effectiveness of the investigated biosensor was also investigated in human serum. The EIS comparative investigations were performed in human serum containing a concentration of 1×10 −12  M (0.5 mIU/mL) hCG and in human serum where the hCG was added. The obtained results indicate that the investigated biosensor is selective for the presence of hCG hormone in the human serum.</abstract><doi>10.1002/elan.202300095</doi></addata></record>
fulltext fulltext
identifier ISSN: 1040-0397
ispartof Electroanalysis (New York, N.Y.), 2023-11, Vol.35 (11)
issn 1040-0397
1521-4109
language eng
recordid cdi_crossref_primary_10_1002_elan_202300095
source Wiley-Blackwell Read & Publish Collection
title An electrochemical biosensor for the determination of hormone Human Chorionic Gonadotropin (hCG) in human serum
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-02T07%3A09%3A45IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-crossref&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=An%20electrochemical%20biosensor%20for%20the%20determination%20of%20hormone%20Human%20Chorionic%20Gonadotropin%20(hCG)%20in%20human%20serum&rft.jtitle=Electroanalysis%20(New%20York,%20N.Y.)&rft.au=Koterwa,%20Adrian&rft.date=2023-11&rft.volume=35&rft.issue=11&rft.issn=1040-0397&rft.eissn=1521-4109&rft_id=info:doi/10.1002/elan.202300095&rft_dat=%3Ccrossref%3E10_1002_elan_202300095%3C/crossref%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c194t-e94b52a9dee55a0fa3fbae7814f52607d8b84fd1a32f8e1038597a783e7522393%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_id=info:pmid/&rfr_iscdi=true