Loading…
An evaluation of paper chromatography for measuring the levels of radiochemical impurities in 99m Tc Medronate Injection
Background : The European Pharmacopoeia specifies Instant Thin‐Layer Chromatography Silica Gel (ITLC‐SG) as the stationary phase for measuring radiochemical impurities in 99m Tc Medronate. The Pall Corporation has stopped manufacturing ITLC‐SG. Aim : To investigate chromatography papers as alternati...
Saved in:
Published in: | Journal of labelled compounds & radiopharmaceuticals 2010-01, Vol.53 (1), p.11-14 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Background
: The European Pharmacopoeia specifies Instant Thin‐Layer Chromatography Silica Gel (ITLC‐SG) as the stationary phase for measuring radiochemical impurities in
99m
Tc Medronate. The Pall Corporation has stopped manufacturing ITLC‐SG.
Aim
: To investigate chromatography papers as alternatives to ITLC‐SG.
Experimental
: The resolution of the main
99m
Tc Medronate and impurity peaks were compared on five Whatman papers (1CHR, 31ET, 3MM, 4CHR and 54SFC) using the pharmacopoeial mobile phases: methyl ethyl ketone for detecting
99m
Tc pertechnetate impurity and sodium acetate solution (136 g/l) for detecting hydrolysed reduced and colloidal
99m
Tc impurities. The effects of three sample spot treatments (wet, hot dried and air dried) on the distribution of the radiochemical impurities were compared. The levels of radiochemical impurities measured in samples of
99m
Tc Medronate were compared.
Results
: The highest resolutions were obtained with 1CHR, 4CHR and 54SFC. Sample spot drying resulted in the measurement of artifactually high levels of
99m
Tc pertechnetate impurity. When analysing
99m
Tc Medronate, erroneously high levels of colloidal and
99m
Tc pertechnetate impurities were measured with 1CHR and 4CHR.
Conclusions
: 54SFC paper is a suitable alternative to ITLC‐SG for measuring radiochemical impurities in
99m
Tc Medronate. The sample spot should be wet when the paper strip is placed in the chromatography tank. Copyright © 2009 John Wiley & Sons, Ltd. |
---|---|
ISSN: | 0362-4803 1099-1344 |
DOI: | 10.1002/jlcr.1693 |