Loading…
A simplified and higher dynamic range version of the J 1 … J 4 method for optical phase measurements
A new and an improved version of the J 1 …J 4 method for optical phase detection, named J 1 –J 3 method, is described, which is more stable to phase drift and simpler for implementation than the original one. The sensor dynamic range is increased while maintaining the accuracy and the minimum detect...
Saved in:
Published in: | Microwave and optical technology letters 2012-02, Vol.54 (2), p.387-392 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A new and an improved version of the J
1
…J
4
method for optical phase detection, named J
1
–J
3
method, is described, which is more stable to phase drift and simpler for implementation than the original one. The sensor dynamic range is increased while maintaining the accuracy and the minimum detectable phase‐shift. The method is experimentally validated by using an electro‐optic intensity modulator. The agreement between theoretical and experimental results, based on the 1/f voltage noise, is demonstrated. © 2011 Wiley Periodicals, Inc. Microwave Opt Technol Lett, 54:387–392, 2012; View this article online at wileyonlinelibrary.com. DOI 10.1002/mop.26533 |
---|---|
ISSN: | 0895-2477 1098-2760 |