Effects of B 0 eddy currents on imaging isocenter shifts in 0.35-T MRI-guided radiotherapy (MR-IGRT) system

The purpose of this study was to measure gantry angle-related eddy currents in a 0.35-T MRI-Linac and determine if B (zeroth order) eddy currents are the primary cause of gantry angle-dependent imaging isocenter shifts vs other potential causes like B inhomogeneities and gradient (first order) eddy...

Full description

Saved in:
Bibliographic Details
Published in:Medical physics (Lancaster) 2021-06, Vol.48 (6), p.2929-2938
Main Authors: Curcuru, Austen N, Lewis, Benjamin C, Kim, Taeho, Yang, Deshan, Michael Gach, H
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The purpose of this study was to measure gantry angle-related eddy currents in a 0.35-T MRI-Linac and determine if B (zeroth order) eddy currents are the primary cause of gantry angle-dependent imaging isocenter shifts vs other potential causes like B inhomogeneities and gradient (first order) eddy currents. For conventional Cartesian acquisitions, B eddy currents can cause imaging isocenter shifts along both phase encode and readout directions. Gradient eddy currents can cause spatial distortion along both the phase encode and readout directions. Center frequency offsets can cause imaging isocenter shifts along the readout direction that vary with readout gradient polarity. MRI-related eddy currents and imaging isocenter shifts were measured on a 0.35-T MRI-Linac at gantry angles from 0° to 330° in increments of 30 . All measurements were made after gradient shimming and center frequency tuning at each planned gantry angle. Eddy current and field homogeneity measurements were conducted using a 24-cm diameter spherical phantom. Gradient and B eddy currents were calculated from the free induction decays (FIDs) resulting from selective excitation of slices located ±5 cm from isocenter. B eddy currents were also calculated from FIDs acquired with nonselective excitation and compared with B eddy current values derived using selective excitation. B inhomogeneities and center frequency offsets were measured by acquiring FIDs with nonselective excitation. Imaging isocenter shifts were measured using a 33x33x10.5 cm uniformity linearity (grid) phantom and a 3D true fast imaging with steady-state precession (TrueFISP) sequence used in MRI-guided radiation therapy. Eddy currents were compared to vendor specifications and correlated with the imaging isocenter shifts. Measurements were conducted before and after the MRI-Linac's waveguide was replaced with an updated design to reduce eddy currents. B eddy currents were highly correlated (r = 0.986, P 
ISSN:0094-2405
2473-4209
DOI:10.1002/mp.14842