Loading…
Reduction of transmitter B 1 inhomogeneity with transmit SENSE slice‐select pulses
Parallel transmitter techniques are a promising approach for reducing transmitter B 1 inhomogeneity due to the potential for adjusting the spatial excitation profile with independent RF pulses. These techniques may be further improved with transmit sensitivity encoding (SENSE) methods because the se...
Saved in:
Published in: | Magnetic resonance in medicine 2007-05, Vol.57 (5), p.842-847 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Parallel transmitter techniques are a promising approach for reducing transmitter
B
1
inhomogeneity due to the potential for adjusting the spatial excitation profile with independent RF pulses. These techniques may be further improved with transmit sensitivity encoding (SENSE) methods because the sensitivity information in pulse design provides an excitation that is inherently compensated for transmitter
B
1
inhomogeneity. This paper presents a proof of this concept using transmit SENSE 3D tailored RF pulses designed for small flip angles. An eight‐channel receiver coil was used to mimic parallel transmission for brain imaging at 3T. The transmit SENSE pulses were based on the fast‐
k
z
design and produced 5‐mm‐thick slices at a flip angle of 30° with only a 4.3‐ms pulse length. It was found that the transmit SENSE pulses produced more homogeneous images than those obtained from the complex sum of images from all receivers excited with a standard RF pulse. Magn Reson Med 57:842–847, 2007. © 2007 Wiley‐Liss, Inc. |
---|---|
ISSN: | 0740-3194 1522-2594 |
DOI: | 10.1002/mrm.21221 |