Loading…

An Improved Argon Microwave-Induced Plasma Ionization Detector for Gas Chromatography

The microwave-induced plasma ionization detector (MIPID) for gas chromatography has been improved. The effect of the structure of MIPID and experimental parameters on its performance were studied with argon as the discharge and carrier gas. The sensitivity, linear dynamic range, noise, and stability...

Full description

Saved in:
Bibliographic Details
Published in:Microchemical journal 1999-06, Vol.62 (2), p.282-290
Main Authors: Jin, Qinhan, Shi, Yuhua, Yu, Aimin, Yang, Wenjun, Peng, Zenghui
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The microwave-induced plasma ionization detector (MIPID) for gas chromatography has been improved. The effect of the structure of MIPID and experimental parameters on its performance were studied with argon as the discharge and carrier gas. The sensitivity, linear dynamic range, noise, and stability of three types of MIPIDs have been compared for various compounds. For most organic compounds, detection limits in the range 10 −10–10 −11g with a linear dynamic range of four to five orders of magnitude were obtained. The influence of air entrainment into the plasma on the analytical performance of the detector was also studied.
ISSN:0026-265X
1095-9149
DOI:10.1006/mchj.1999.1732