Loading…
Application of x-ray fluorescence to the study of iodine distribution and content in the thyroid
We have been developing an X-ray fluorescence system designed to determine iodine distribution and content in the thyroid. The X-ray fluorescence unit is composed of a 80 KV X-ray excitation beam and a Si(Li) semi-conductor detector. The angle between the X-ray beam and the axis of the detector is 2...
Saved in:
Published in: | European Journal of Nuclear Medicine 1981-01, Vol.6 (9), p.407-410 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We have been developing an X-ray fluorescence system designed to determine iodine distribution and content in the thyroid. The X-ray fluorescence unit is composed of a 80 KV X-ray excitation beam and a Si(Li) semi-conductor detector. The angle between the X-ray beam and the axis of the detector is 24 degree. Two single channel analyzers (SCA) are used, the first one corresponding to the K alpha peak and the other to the background. The SCA signals are digitised so that the two corresponding images may be processed using a computer. After subtracting the background, an image is obtained that represents the iodine distribution alone, and the iodine content of the whole thyroid or a part of it can be determined with a reproducibility of +/- 5%. A good correlation (r=0.98) between iodine chemical analysis and iodine fluorescence analysis was obtained for 13 patients. |
---|---|
ISSN: | 0340-6997 1619-7089 |
DOI: | 10.1007/BF00266431 |