Fabrication and Determination of an Aluminium Fixed Point Cell as a Secondary Standard in Thermocouple Calibration
A new open-type aluminium (Al) fixed-point cell has been fabricated in Puslit Metrologi-LIPI. The cell was successfully characterized by direct comparison with a commercial Al cell. The cell plateau lasted for four hours with a flatness of 0.014 °C, measured as the difference between the beginning a...
Saved in:
Published in: | MĀPAN : journal of Metrology Society of India 2016-03, Vol.31 (1), p.57-60 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A new open-type aluminium (Al) fixed-point cell has been fabricated in Puslit Metrologi-LIPI. The cell was successfully characterized by direct comparison with a commercial Al cell. The cell plateau lasted for four hours with a flatness of 0.014 °C, measured as the difference between the beginning and the end of the plateau using a type S thermocouple. The reproducibility of realizations is 0.04 °C; the immersion profile of the cell was measured for 30 mm depth, from the bottom of the cell, and the variation was found to be 0.02 °C. A good agreement within 0.16 °C was measured at the freezing point. Furthermore, the cell improves the entry in Appendix C, calibration and measurement capability (CMC) by 0.24 °C, allowing the cell to be included as a secondary standard of the international temperature scale of 1990 realization in Puslit Metrologi-LIPI. |
---|---|