Rapid detection of l-glutamic acid using a series-electrode piezoelectric quartz crystal sensor
Two methods are described for the determination of l-glutamic acid using a series-electrode piezoelectric quartz crystal sensor. The effect of formaldehyde solution on the frequency shift is discussed. For the calibration curve method, in formaldehyde medium, the detection limit is 3.8 · 10 −5 M, th...
Saved in:
Published in: | Analytica chimica acta 1995-09, Vol.313 (3), p.229-236 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Two methods are described for the determination of
l-glutamic acid using a series-electrode piezoelectric quartz crystal sensor. The effect of formaldehyde solution on the frequency shift is discussed. For the calibration curve method, in formaldehyde medium, the detection limit is 3.8 · 10
−5 M, the recovery is 100.4%, the relative standard deviation is 3.32% (
n = 5) and the frequency shift is linear up to 8 · 10
−4 M; in aqueous medium, the above values are 2.0 · 10
−5 M, 99.1%, 3.01% (
n = 6) and 1.1 · 10
−3 M, respectively. The frequencimetric titration method is also described and the lowest titratable concentration of
l-glutamic acid is 7.0 · 10
−5 M. The detecting results for tablet samples show good agreement with the values determined by the pharmaceutical method. |
---|---|
ISSN: | 0003-2670 1873-4324 |
DOI: | 10.1016/0003-2670(95)00234-Q |