On-line determination of the ash content of coal using a “Siroash” gauge based on the transmission of low and high energy γ-rays

A gauge based on measurements of the transmission through coal of low and high energy γ-rays has been developed for the on-line determination of the ash content. The SIROASH (C SIRO ASH) gauge has been tested on the final product line of a coal washery at Gregory, Queensland, Australia. The r.m.s. d...

Full description

Saved in:
Bibliographic Details
Published in:The International journal of applied radiation and isotopes 1983, Vol.34 (1), p.63-69
Main Authors: Fookes, R.A., Gravitis, V.L., Watt, J.S., Hartley, P.E., Campbell, C.E., Howells, E., McLennan, T., Millen, M.J.
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A gauge based on measurements of the transmission through coal of low and high energy γ-rays has been developed for the on-line determination of the ash content. The SIROASH (C SIRO ASH) gauge has been tested on the final product line of a coal washery at Gregory, Queensland, Australia. The r.m.s. difference between ash determined on-line and ash determined by chemical assay of coal sampled from the conveyor was 0.45 wt 0 0. Chemical assay and sampling errors, estimated to be between 0.2 and 0.3 wt 0 0 ash, contribute to this difference. The SIROASH gauge measures coal directly on the conveyor, and can be calibrated using standard samples by moving it beyond the edge of the conveyor belt.