Improvements of analytical results of thermal ionization mass spectrometry through electrolytic sample deposition

A method of sample deposition based on electrodeposition in a dimethyl formamide media is proposed. The apparatus and procedure described lead to centered and reproducible deposits on three filaments assemblies. The same procedure is used for all the actinides. The electrodeposition yields range fro...

Full description

Saved in:
Bibliographic Details
Published in:International journal of mass spectrometry and ion physics 1983, Vol.48, p.393-396
Main Authors: Bergey, C., Cesario, J., Deniaud, S.
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A method of sample deposition based on electrodeposition in a dimethyl formamide media is proposed. The apparatus and procedure described lead to centered and reproducible deposits on three filaments assemblies. The same procedure is used for all the actinides. The electrodeposition yields range from 40 to 90 %. Mass spectrometry results demonstrate improved precision and better sensitivity. The simplicity of manipulations allows easy automation.