Gravimetric calibration of standard gases

Permeation rates of specific gases (e.g. SO 2, NO 2) in the range of 10 −9 g min −1 were directly measured using a high performance microbalance. The gases were trapped in small permeation tubes with total weights of about 2.5 g. By controlling the microbalance's temperature the measurements we...

Full description

Saved in:
Bibliographic Details
Published in:Thermochimica acta 1986-07, Vol.103 (1), p.57-62
Main Authors: Beltz, N., Jaeschke, W., Meixner, F.X.
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Permeation rates of specific gases (e.g. SO 2, NO 2) in the range of 10 −9 g min −1 were directly measured using a high performance microbalance. The gases were trapped in small permeation tubes with total weights of about 2.5 g. By controlling the microbalance's temperature the measurements were done at set temperatures between 25°C and 60°C to obtain a reference graph (permeation rate vs. temperature) for each gas. The microbalance was connected to a metered flow of a clean-air device. With adjustable flow rates up to 50 1 min −1 the permeating gas could be diluted down to volume mixing ratios in the 10 −9 and even the 10 −12 range.
ISSN:0040-6031
1872-762X