Test measurements and analysis of errors for a new equipment for the determination of excess-heat data

In this paper the systematic application of error analysis, error propagation and sensitivity calculations for the improvement of a calorimeter is shown. Adaptions were made on a SETARAM C-80 heat flux calorimeter to perform measurements of excess-heat data above the boiling point of the components....

Full description

Saved in:
Bibliographic Details
Published in:Thermochimica acta 1991-09, Vol.187, p.95-112
Main Authors: Huemer, H, Platzer, E, Rehak, K
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:In this paper the systematic application of error analysis, error propagation and sensitivity calculations for the improvement of a calorimeter is shown. Adaptions were made on a SETARAM C-80 heat flux calorimeter to perform measurements of excess-heat data above the boiling point of the components. Test measurements were performed for the system n-hexane - c-hexane at 25, 40, 60 and 80 °C and for glycerol - water at 50 and 70 °C. The analysis of these results, error and sensitivity calculations show the sources of error and permit improvement and tuning of the equipment, which is demonstrated by improved results.