Test measurements and analysis of errors for a new equipment for the determination of excess-heat data
In this paper the systematic application of error analysis, error propagation and sensitivity calculations for the improvement of a calorimeter is shown. Adaptions were made on a SETARAM C-80 heat flux calorimeter to perform measurements of excess-heat data above the boiling point of the components....
Saved in:
Published in: | Thermochimica acta 1991-09, Vol.187, p.95-112 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | In this paper the systematic application of error analysis, error propagation and sensitivity calculations for the improvement of a calorimeter is shown. Adaptions were made on a SETARAM C-80 heat flux calorimeter to perform measurements of excess-heat data above the boiling point of the components. Test measurements were performed for the system n-hexane - c-hexane at 25, 40, 60 and 80 °C and for glycerol - water at 50 and 70 °C. The analysis of these results, error and sensitivity calculations show the sources of error and permit improvement and tuning of the equipment, which is demonstrated by improved results. |
---|---|
ISSN: | 0040-6031 1872-762X |
DOI: | 10.1016/0040-6031(91)87185-Y |