Absolute brightness measurement of X-ray emission from laser-produced plasma using thermoluminescent dosemeters
Thermoluminescent dosemeters of LiF:Mg,Ti, LiF:Mg,Cu,P and CaF2:Dy were used as secondary standards in calibration measurements of X-ray emission from a laser-produced aluminium plasma. Relative sensitivities of the thermoluminescent dosemeters were found to be independent of dose rate over a wide d...
Saved in:
Published in: | Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 1996-01, Vol.368 (2), p.484-487 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Thermoluminescent dosemeters of LiF:Mg,Ti, LiF:Mg,Cu,P and CaF2:Dy were used as secondary standards in calibration measurements of X-ray emission from a laser-produced aluminium plasma. Relative sensitivities of the thermoluminescent dosemeters were found to be independent of dose rate over a wide dose rate region. The photon detection efficiency of thermoluminescent dosemeters is compared to that of semiconductor Si photodiodes. The thermoluminescent dosemeters are presented as a powerful tool for testing of capability of a semiconductor photodiode detector of X-ray pulse energy emitted from the laser-produced plasma. |
---|---|