At the Department of Physics of the University of Bologna a new intensified linear array detector is under development. The core of the system is a digital intensified CCD camera, the electron bombarded charge coupled device (EBCCD). The main innovation is a coherent rectangular-to-linear fiber opti...
Saved in:
Published in: | Nuclear instruments & methods in physics research. Section B, Beam interactions with materials and atoms Beam interactions with materials and atoms, 2004, Vol.213, p.227-230 |
---|---|
Main Authors: | , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | At the Department of Physics of the University of Bologna a new intensified linear array detector is under development. The core of the system is a digital intensified CCD camera, the electron bombarded charge coupled device (EBCCD). The main innovation is a coherent rectangular-to-linear fiber optics adapter coupling the 1 in. diameter photocathode of the camera with a linear 129 mm
×
1.45 mm strip of Gd
2O
2S:Tb. In this way a high spatial resolution over an extended length is obtained. The detector works as an X-ray scanner by means of a high-precision translation mechanical device to inspect a 13 cm
×
18 cm area. A complete characterisation of the system has been made in terms of linearity, dynamic range, modulation transfer function (MTF), noise power spectrum (NPS) and detective quantum efficiency (DQE). At last, radiographic tests on a set of samples have been made and will be presented. |
---|---|