Loading…

Commissioning Results of the BTUC-PGM beamline

The CRG beamline was commissioned in June 2000. This paper presents the results of the first resolution and flux measurements. The monochromator resolution was evaluated using gas cell absorbtion measurements for argon, nitrogen and neon. The results of the nitrogen measurements demonstrate a resolv...

Full description

Saved in:
Bibliographic Details
Published in:Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2001-07, Vol.467, p.470-473
Main Authors: Batchelor, D.R., Follath, R., Schmeißer, D.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The CRG beamline was commissioned in June 2000. This paper presents the results of the first resolution and flux measurements. The monochromator resolution was evaluated using gas cell absorbtion measurements for argon, nitrogen and neon. The results of the nitrogen measurements demonstrate a resolving power of 10,000. The monochromator flux measurements were made using a GaAs diode and show that the beamline can deliver 10 13 photons s −1 at 100 eV for a 140 μm exit slit.
ISSN:0168-9002
1872-9576
DOI:10.1016/S0168-9002(01)00362-X