Performance study of liquid xenon detector for PET

The coincidence time resolution and the intrinsic position resolution along the electron drift direction of a liquid xenon multiwire ionization chamber for PET were measured, being, respectively, 1.5 ns (fwhm) and 0.5 mm (fwhm). Data on the low temperature performance of a photomultiplier tube opera...

Full description

Saved in:
Bibliographic Details
Published in:Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 1997-06, Vol.392 (1), p.427-432
Main Authors: Chepel, V, Lopes, M.I, Kuchenkov, A, Ferreira Marques, R, Policarpo, A.J.P.L
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The coincidence time resolution and the intrinsic position resolution along the electron drift direction of a liquid xenon multiwire ionization chamber for PET were measured, being, respectively, 1.5 ns (fwhm) and 0.5 mm (fwhm). Data on the low temperature performance of a photomultiplier tube operating on the VUV are presented.