A simple circuit for background reduction in low level tritium activity measurement using liquid-scintillation counting
The paper describes a new method for reduction of background counts in Liquid Scintillation Counting (LSC) using a simple circuit. With this circuit it was possible to reduce background counts in low level Tritium Counting to 8 cpm from an initial value of 18 cpm with a marginal degradation in detec...
Saved in:
Published in: | Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 1998-07, Vol.411 (1), p.188-190 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The paper describes a new method for reduction of background counts in Liquid Scintillation Counting (LSC) using a simple circuit. With this circuit it was possible to reduce background counts in low level Tritium Counting to 8
cpm from an initial value of 18
cpm with a marginal degradation in detection efficiency. Use of the circuit can reduce the cost of passive shielding, which is generally used to reduce background counts to a low level. |
---|---|