Performance evaluation of MRI-based PAGAT polymer gel dosimeter in an inhomogeneous phantom using EGSnrc code on a Co-60 machine
Relative isodose curves were obtained using PAGAT gel dosimeter on homogeneous and inhomogeneous phantoms. Distance-to-agreement (DTA) was calculated between simulated and measured values for both the homogeneous and inhomogeneous phantoms. All DTAs except one passed the acceptance criterion (±5 dos...
Saved in:
Published in: | Applied radiation and isotopes 2009, Vol.67 (1), p.186-191 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Relative isodose curves were obtained using PAGAT gel dosimeter on homogeneous and inhomogeneous phantoms. Distance-to-agreement (DTA) was calculated between simulated and measured values for both the homogeneous and inhomogeneous phantoms. All DTAs except one passed the acceptance criterion (±5 dose variation for selected isodose levels). Results of this study also showed the ability of the Monte Carlo modeling to provide accurate dosimetry, and revealed that the dose response of PAGAT polymer gel is dependent on the method of fabrication. |
---|---|