Loading…

Standardization of 106 Ru/Rh by live-timed anticoincidence counting and gamma emission determination

The absolute activity standardization measurement system of radionuclide by live-timed anticoincidence counting was implemented at LNMRI in 2008 to reduce the effects of some correction factors on the determination of activity with coincidence counting technique used for decades in the laboratory, f...

Full description

Saved in:
Bibliographic Details
Published in:Applied radiation and isotopes 2017-04, Vol.122, p.37-40
Main Authors: da Silva, C J, Rezende, E A, Poledna, R, Tauhata, L, Iwahara, A, Lopes, R T
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The absolute activity standardization measurement system of radionuclide by live-timed anticoincidence counting was implemented at LNMRI in 2008 to reduce the effects of some correction factors on the determination of activity with coincidence counting technique used for decades in the laboratory, for example, the corrections of dead time and resolution. With the live-timed anticoincidence system, the variety of radionuclides that can be calibrated by LNMRI was increased in relation to the type of decay. The objective of this study was to standardize the Ru activity, determine gamma emission probabilities by spectrometric method for some energies, and estimate measurement uncertainties.
ISSN:0969-8043
1872-9800
DOI:10.1016/j.apradiso.2017.01.007