Loading…
Fractures of Modern High Nitrogen Stainless Steel Cemented Stems
Abstract We present 14 cases of fracture of modern, high-nitrogen, stainless steel stems. Our clinical and radiological data suggest that heavy patients with small stems and poor proximal support are at risk for fracturing their implants. “Champagne-glass” canals can lead to the use of smaller stems...
Saved in:
Published in: | The Journal of arthroplasty 2008-02, Vol.23 (2), p.188-196 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Abstract We present 14 cases of fracture of modern, high-nitrogen, stainless steel stems. Our clinical and radiological data suggest that heavy patients with small stems and poor proximal support are at risk for fracturing their implants. “Champagne-glass” canals can lead to the use of smaller stems often placed in varus, which can lead to cantilever bending and fatigue failure in the distal half of the stem. Metallurgical assessment of the retrieved high-nitrogen, stainless steel stems reveals microstructural inconsistencies that may contribute to their failure. Based on our findings, careful consideration and attention to technique is required when using stainless steel stems in patients with high body mass index or high weight. Technique is particularly important in femurs with champagne-glass canals. |
---|---|
ISSN: | 0883-5403 1532-8406 |
DOI: | 10.1016/j.arth.2006.12.051 |