Risk reduction benefits of cryogen-free refrigeration

High Purity Germanium (HPGe) gamma-ray detectors are used at Los Alamos National Laboratory's Plutonium Facility (TA-55) for nondestructive assay. Liquid nitrogen (LN2), a cryogen, is commonly used to cool these detectors. Cryogen use is associated with several health risks and operational prob...

Full description

Saved in:
Bibliographic Details
Published in:ACS Chemical Health & Safety. 2012-05, Vol.19 (3), p.29-33
Main Authors: Cournoyer, Michael E., Pecos, James M., Chunglo, Steve D., Bonner, Charles, Foster, Lynn A., Maez, Rudy J., Punjak, Wayne A., Wilburn, Dianne W.
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:High Purity Germanium (HPGe) gamma-ray detectors are used at Los Alamos National Laboratory's Plutonium Facility (TA-55) for nondestructive assay. Liquid nitrogen (LN2), a cryogen, is commonly used to cool these detectors. Cryogen use is associated with several health risks and operational problems. This has prompted the development of cryogen-free refrigeration. Although the efficacy of the coolers has been proven for use with HPGe detectors, the health and safety aspect implications have not been studied. Both LN2 and mechanical coolers were examined and their application effectiveness on health and safety aspects in HPGe gamma-ray detector use at TA-55 was compared. Overall, cryogen-free refrigeration is an engineering equivalent to LN2 systems used in gamma spectroscopy and eliminates the health and physical hazards associated with LN2 systems without adding hazards.
ISSN:1871-5532
DOI:10.1016/j.jchas.2011.12.005