Spectral broadening by spatial effects in X-ray microcalorimeters with superconducting absorber and NTD-Ge thermal sensor
We discuss how the sensitivity to the position of absorption of the photons can affect the spectral resolution of X-ray microcalorimeters with superconducting Sn absorber and NTD-Ge thermal sensor. Data obtained through the computer simulations are reported for microcalorimeters with different geome...
Saved in:
Published in: | Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2004-03, Vol.520 (1-3), p.216-219 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We discuss how the sensitivity to the position of absorption of the photons can affect the spectral resolution of X-ray microcalorimeters with superconducting Sn absorber and NTD-Ge thermal sensor. Data obtained through the computer simulations are reported for microcalorimeters with different geometries and phonon transmission probabilities at the interface between the absorber and thermistor. |
---|---|
ISSN: | 0168-9002 1872-9576 |
DOI: | 10.1016/j.nima.2003.11.232 |