Synthetic single crystal diamond diodes for radiotherapy dosimetry
Synthetic single crystal diamonds in a p-type/intrinsic/metal structure were tested as dosimeters for radiotherapy. The devices have been analyzed by using 6 and 10 MV Bremsstrahlung X-ray beams and electron beams from 6 MeV up to 18 MeV from a CLINAC DHX Varian accelerator. All measurements have be...
Saved in:
Published in: | Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment Accelerators, spectrometers, detectors and associated equipment, 2008-09, Vol.594 (2), p.273-277 |
---|---|
Main Authors: | , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Synthetic single crystal diamonds in a p-type/intrinsic/metal structure were tested as dosimeters for radiotherapy. The devices have been analyzed by using 6 and 10
MV Bremsstrahlung X-ray beams and electron beams from 6
MeV up to 18
MeV from a CLINAC DHX Varian accelerator. All measurements have been performed in a water phantom and ionization chambers were used for calibration and comparison. The dosimeters were operated in photovoltaic regime with no external bias voltage applied. A few Gy pre-irradiation was performed in order to stabilize the device output, resulting in fluctuations sensitivity below ±0.5%. No dose rate dependence of the detector response was observed. Very good reproducibility and linearity were obtained as well. |
---|---|
ISSN: | 0168-9002 1872-9576 |
DOI: | 10.1016/j.nima.2008.06.028 |