Amperometric Monitoring of the Neuropeptide “Substance-P” Level in Biological Fluids
Presented here is a capillary enzyme immunoassay constructed on a foil-based microfluidic chip for rapid determination of Substance-P (SP) level in human plasma, serum and saliva. Using a competitive ELISA implemented into the capillary, the amperometrically detected SP by flow injection analysis wa...
Saved in:
Published in: | Procedia chemistry 2009-09, Vol.1 (1), p.1275-1278 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Presented here is a capillary enzyme immunoassay constructed on a foil-based microfluidic chip for rapid determination of Substance-P (SP) level in human plasma, serum and saliva. Using a competitive ELISA implemented into the capillary, the amperometrically detected SP by flow injection analysis was related to the percentage amount between SP standard (or sample) and SP-labelled glucose oxidase conjugate bound to the capture antibody. A detection limit of 10 pg/ml was obtained. The assay allows fast reading of diagnostically relevant levels. Producing reliable results comparable to microplate format, the immunochip sensor may have a significant impact in point of care diagnostics. |
---|---|