Loading…
An Optoelectrical, Standard CMOS-based Active Catheter Tracking System for MRI
A fully optical active catheter tracking system compatible with 3T M RI environment is presented. It replaces conducting cables with optical fibers to reduce RF-induced heating problem. Proposed system consists of a M EM S-based microstructure array and an IC driving it. The IC houses an RF receiver...
Saved in:
Main Authors: | , , |
---|---|
Format: | Conference Proceeding |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A fully optical active catheter tracking system compatible with 3T M RI environment is presented. It replaces conducting cables with optical fibers to reduce RF-induced heating problem. Proposed system consists of a M EM S-based microstructure array and an IC driving it. The IC houses an RF receiver block and an optoelectrical power supply. A prototype IC was fabricated in UM C 0.18μm CM OS process. M easurements indicate that the supply unit is able to provide 2.18mA at 1.2V supply, when a laser beam of 80mW power at its source is applied to IC. The system is operational at laser source power levels above 40mW. |
---|---|
ISSN: | 1877-7058 1877-7058 |
DOI: | 10.1016/j.proeng.2014.11.725 |