The dose response of three LiF TLDs: standard LiF:Mg,Ti (denoted MTS), high-sensitive LiF:Mg,Cu,P (MCP) and a recently developed in Kraków version of LiF:Mg,Ti with modified activator composition (MTT) and increased high-LET response was measured. The TLDs have been exposed to 60Co gamma-rays, up to...
Saved in:
Published in: | Radiation measurements 2007-04, Vol.42 (4), p.582-585 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The dose response of three LiF TLDs: standard LiF:Mg,Ti (denoted MTS), high-sensitive LiF:Mg,Cu,P (MCP) and a recently developed in Kraków version of LiF:Mg,Ti with modified activator composition (MTT) and increased high-LET response was measured. The TLDs have been exposed to
60Co gamma-rays, up to dose of 10
000
Gy, i.e. beyond saturation dose of the main dosimetric peaks, which corresponds to ca. 1000
Gy. The measured glow-curves were deconvolved into separate peaks with first order kinetic function (using self-developed GlowFit software). The dose response of the main peaks was found to be supralinear for MTS and sublinear for MCP detectors, as expected. The dose response of MTT was found to be even more supralinear than that of MTS. An interesting effect has been observed with regard to glow-curve shape of MCP detectors. Up to a dose of 1
kGy it remains practically unchanged, while for higher doses a strong growth of high-temperature peaks is observed. In the same dose region a decrease of the main peak of MCP with increasing dose is observed, unlike LiF:Mg,Ti detectors. |
---|---|
ISSN: | 1350-4487 1879-0925 |
DOI: | 10.1016/j.radmeas.2007.01.077 |