Loading…
Development of digital imaging device based on the gas electron multiplier
We fabricated a gas electron multiplier (GEM) for application to digital imaging devices such as X-ray or γ -ray detectors. Sixty-four square ( 3 × 3 mm ) copper pads were used to collect avalanched electron charge clusters from the GEM chamber. A 128-channel charge sensitive preamplifier, which als...
Saved in:
Published in: | Radiation measurements 2007-11, Vol.42 (10), p.1661-1665 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We fabricated a gas electron multiplier (GEM) for application to digital imaging devices such as X-ray or
γ
-ray detectors. Sixty-four square (
3
×
3
mm
) copper pads were used to collect avalanched electron charge clusters from the GEM chamber. A 128-channel charge sensitive preamplifier, which also includes sample/hold amplifiers and a multiplexer, was used as the front end electronics of the readout circuits for data acquisition. With this system, we successfully took a two-dimensional digital image under the irradiation of the
55Fe radiation source. |
---|---|
ISSN: | 1350-4487 1879-0925 |
DOI: | 10.1016/j.radmeas.2007.09.009 |