Loading…

A novel gas-filled detector for pulsed X-ray sources

A novel gas-filled detector developed for use in wavelength-dispersive X-ray spectrometers operated with pulsed X-ray sources is described. The design of the detector is based on the principle used in drift chambers. The effects of geometric parameters and gas composition on the detection characteri...

Full description

Saved in:
Bibliographic Details
Published in:Spectrochimica acta. Part B: Atomic spectroscopy 2004-08, Vol.59 (8), p.1329-1334
Main Authors: Beaven, Peter A., Marmotti, Mauro, Kampmann, Reinhard, Knoth, Joachim, Schwenke, Heinrich
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A novel gas-filled detector developed for use in wavelength-dispersive X-ray spectrometers operated with pulsed X-ray sources is described. The design of the detector is based on the principle used in drift chambers. The effects of geometric parameters and gas composition on the detection characteristics have been calculated using Monte-Carlo techniques. According to these calculations the multichamber drift detector we have constructed should be capable of acquiring one to approximately 300 low-energy (1.5 keV) photons per burst.
ISSN:0584-8547
1873-3565
DOI:10.1016/j.sab.2004.05.018