Voltammetric Quantification of 4‑Nitrophenol Using a Multivariate Optimized Plated Bismuth Film Electrode

In this study the multivariate optimization of bismuth film deposition (BiFE) on glassy carbon electrode, together with square wave voltammetry parameters for nitrophenol quantification, is described for the first time. For this purpose, response surface methodology and desirability function were ap...

Full description

Saved in:
Bibliographic Details
Published in:Industrial & engineering chemistry research 2019-07, Vol.58 (27), p.12411-12418
Main Authors: Nuñez, Rodrigo N, Moreno Betancourth, Jessica, Ortiz, Patricia I, Pfaffen, Valeria
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:In this study the multivariate optimization of bismuth film deposition (BiFE) on glassy carbon electrode, together with square wave voltammetry parameters for nitrophenol quantification, is described for the first time. For this purpose, response surface methodology and desirability function were applied to a Box–Behnken design. For the BiFE deposition procedure, the relationship between the reduction current peak (i p) and standard deviation values with time, potential deposition, and Bi­(III) concentration were analyzed; meanwhile, for frequency, step, and amplitude only the relationship with i p was considered. The pH influence in nitrophenol determination was carried on by univariate analysis. After parameters optimization a limit of detection of 3.4 × 10–6 mol L–1 was observed.
ISSN:0888-5885
1520-5045
DOI:10.1021/acs.iecr.9b01700