Determination of strontium 90 in environmental samples by triple quadrupole ICP-MS and its application to Fukushima soil samples
The determination of 90 Sr by ICP-MS is capable of providing a high sample throughput compared with radiometric analytical methods. Although the isobaric interference caused by 90 Zr ions can be reduced by reaction with oxygen, the high background caused by stable 88 Sr has made it difficult to dete...
Saved in:
Published in: | Journal of analytical atomic spectrometry 2018-01, Vol.33 (6), p.181-185 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The determination of
90
Sr by ICP-MS is capable of providing a high sample throughput compared with radiometric analytical methods. Although the isobaric interference caused by
90
Zr ions can be reduced by reaction with oxygen, the high background caused by stable
88
Sr has made it difficult to determine
90
Sr by conventional ICP-MS. In this study, a rapid and sensitive method was developed for the determination of
90
Sr using triple quadrupole ICP-MS. The MS/MS configuration improves the abundance sensitivity by more than two orders of magnitude (5 × 10
−12
), and the detection limit for
90
Sr in 1 g environmental samples was 0.6 Bq kg
−1
. The applicability of the method to the measurement of
90
Sr in environmental samples was demonstrated by analysis of certified reference materials (soil: IAEA 375; plant: IAEA 330, IAEA 373) contaminated by the Chernobyl accident. The measured
90
Sr radioactivities in the samples were consistent with values determined by radiometric analysis, within analytical error. The measured
90
Sr concentration in IAEA 372 was 610 ± 40 Bq kg
−1
(SD,
n
= 3) (Decay correction was made to December 1, 2016). This method provides a powerful tool for the investigation of radiostrontium contaminations in the Fukushima area and elsewhere.
The determination of
90
Sr by triple quadrupole ICP-MS is capable of providing a high sample throughput compared with radiometric analytical methods. |
---|---|
ISSN: | 0267-9477 1364-5544 |
DOI: | 10.1039/c8ja00017d |