Loading…
Ultratrace arsenic determination through hydride trapping on oxidized multiwall carbon nanotubes coupled to electrothermal atomic absorption spectrometry
Arsenic determination in natural waters is an issue of current research. This article reports a novel hydride generation (HG) approach developed for As determination with electrothermal atomic absorption spectrometry (ETAAS) detection. The HG process was interfaced with ETAAS through hydride trappin...
Saved in:
Published in: | Journal of analytical atomic spectrometry 2013-01, Vol.28 (6), p.916-922 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Arsenic determination in natural waters is an issue of current research. This article reports a novel hydride generation (HG) approach developed for As determination with electrothermal atomic absorption spectrometry (ETAAS) detection. The HG process was interfaced with ETAAS through hydride trapping onto a carbon nanotubes microcolumn. To this end a homemade gasliquid separator was used, allowing arsine formation and its flow throughout the CNT microcolumn. The retention process involved thus a solid phase extraction from the gas phase to the solid support. Once arsine generation was completed, the elution was carried out with nitric acid directly onto the dosing hole of the graphite furnace. Outstanding sensitivity with detection limit of 1 ng L
1
, quantification limit of 5 ng L
1
and the characteristic mass, 5.8 0.4 pg could be achieved. A satisfactory correlation between concentration of As and absorbance (
R
= 0.9993) from the limit of quantification up to 500 ng L
1
, with a relative standard deviation of 6.3% were obtained. A sensitive enhancement factor of 38 was reached when 2 mL of sample were processed and 50 L of HNO
3
were used as eluent. The system was successfully applied to the analysis of a standard reference material, QC LL2 metals in natural waters. In addition tap water analysis provided an As concentration of 0.29 0.03 g L
1
.
Schematic diagram of the simple manifold employed for As hydride trapping on oxidized multiwall carbon nanotubes with analytical purposes. |
---|---|
ISSN: | 0267-9477 1364-5544 |
DOI: | 10.1039/c3ja30385c |