Loading…
Evaluation of neutron spectra and dose equivalent from a Varian 2100C/D Medical Linear Accelerator: Monte Carlo simulation and a literature review
In this study was carried out a review according to experimental and Monte Carlo studies in the literature on the neutron production from 18 MV, Varian 2100C/D linac. The effects of these neutrons were investigated on the total fluence, the energy spectra, and the dose equivalent. These factors were...
Saved in:
Published in: | Radioprotection 2021-04, Vol.56 (2), p.93-101 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | In this study was carried out a review according to experimental and Monte Carlo studies in the literature on the neutron production from 18 MV, Varian 2100C/D linac. The effects of these neutrons were investigated on the total fluence, the energy spectra, and the dose equivalent. These factors were calculated as a function of depth and the radiation field size by simulation of linac head using of MCNPX2.6.0 code. The neutron strength was found equal to 1.23 × 10
12
nGy
−1
.The results showed that with increasing the field size from 5 × 5 to 40 × 40 cm
2
, the neutron fluence and dose equivalent in the water phantom rose to the maximum value for 25 × 25 cm
2
field (3.05 × 10
7
ncm
−2
Gy
−1
and 3.14 mSvGy
−1
respectively) and then decreased with increasing the field size. According to the results, the magnetite-steel, ordinary, and limonite-steel concrete walls significantly increased the neutron dose equivalent for about 27.4%, 17.2%, and 13.5%, respectively. |
---|---|
ISSN: | 0033-8451 1769-700X |
DOI: | 10.1051/radiopro/2021002 |