Loading…

Noise resolution of RuO2-based resistance thermometers

Low-frequency noise was measured for RuO2-based thick film resistors at liquid helium temperatures down to 0.36 K. The 1∕f-type spectrum and squared voltage dependence of power spectral density observed at low voltages attribute the noise as coming from equilibrium resistance fluctuations. Measureme...

Full description

Saved in:
Bibliographic Details
Published in:Review of scientific instruments 2005-01, Vol.76 (1)
Main Authors: Ptak, Piotr, Kolek, Andrzej, Zawislak, Zbigniew, Stadler, Adam W., Mleczko, Krzysztof
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Low-frequency noise was measured for RuO2-based thick film resistors at liquid helium temperatures down to 0.36 K. The 1∕f-type spectrum and squared voltage dependence of power spectral density observed at low voltages attribute the noise as coming from equilibrium resistance fluctuations. Measurements carried out at different temperatures show that the magnitude of noise intensity (index) increases significantly as temperature goes down. Due to this fact, the resolution of RuO2 thermometers increases above the instrument resolution. The quantity which describes a sensor resolution is defined and calculated for RuO2 thick film sensor. Some remarks on measurement strategy and sensor optimization are supplied.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.1834704