Loading…
First proof-of-principle of inorganic perovskites clinical radiotherapy dosimeters
Inorganic CsPbBr3 perovskite devices have been manufactured and tested as dosimeters under both conventional and Intensity Modulated Radiotherapy (IMRT) X-ray beams. Samples showed a very good linear dependence of the collected charge/current on dose/dose rates in the range of 0.1–5.0 Gy/0.1–4.0 Gy/...
Saved in:
Published in: | APL materials 2019-05, Vol.7 (5), p.051101-051101-6 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Inorganic CsPbBr3 perovskite devices have been manufactured and tested as dosimeters under both conventional and Intensity Modulated Radiotherapy (IMRT) X-ray beams. Samples showed a very good linear dependence of the collected charge/current on dose/dose rates in the range of 0.1–5.0 Gy/0.1–4.0 Gy/min of interest for clinical applications. A device sensitivity of about 70 nC/Gy mm3 compares favorably with other solid-state dosimeters. The first verification of an IMRT dose profile of a prostate cancer treatment, performed by moving the perovskite device on a 10 cm-long profile with a 0.5 mm pitch, showed agreement within 5% with the dose distribution required by the treatment planning system. |
---|---|
ISSN: | 2166-532X 2166-532X |
DOI: | 10.1063/1.5083810 |