Study on Hydrogen Ion-Selective Solid Contact Electrodes Based on Decamethylcyclopentasiloxane
Hydrogen ion-selective solid contact electrode based on decamethylcyclopentasiloxane (DMCS) as ionophore was fabricated. The membrane solution was prepared by mixing DMCS, polyvinyl chloride (PVC), potassium tetrakis p-chlorophenyl borate (KTpClPB) and various plasticizers. The best performance was...
Saved in:
Published in: | Analytical letters 2011-08, Vol.44 (12), p.2138-2149 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Hydrogen ion-selective solid contact electrode based on decamethylcyclopentasiloxane (DMCS) as ionophore was fabricated. The membrane solution was prepared by mixing DMCS, polyvinyl chloride (PVC), potassium tetrakis p-chlorophenyl borate (KTpClPB) and various plasticizers. The best performance was obtained with the sensor based on NPOE (o-nitrophenyl octyl ether) and the conducting polymer layer of poly(pyrrole), doped with NaClO
4
. The electrode exhibited Nernstian-response in the range of pH 1.9-9.8 with a slope of 57.6 ± 0.2 mV per decade and fast response time within 15 s. This electrode showed good selectivity and was successfully used as an indicator electrode in the potentiometric titration. |
---|---|
ISSN: | 0003-2719 1532-236X |
DOI: | 10.1080/00032719.2010.546023 |