Experimental model of occluded biliary metal stent recanalization using irreversible electroporation via a tubular catheter

To demonstrate the feasibility of irreversible electroporation (IRE) for treating biliary metal stent occlusion in an experimental liver model. IRE was performed using an expandable tubular IRE-catheter placed in nitinol stents in the porcine liver. A 3-electrode IRE-catheter was connected to an IRE...

Full description

Saved in:
Bibliographic Details
Published in:International journal of hyperthermia 2021-01, Vol.38 (1), p.393-401
Main Authors: Rohan, T., Andrasina, T., Juza, T., Matkulcik, P., Červinka, D., Svobodova, I., Novotná, V., Bernard, V., Valek, V., Goldberg, S. Nahum
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c479t-c1973a6da200f0ad06e751169199f3ae13344065545b5f8d35fb087150a4d3303
cites cdi_FETCH-LOGICAL-c479t-c1973a6da200f0ad06e751169199f3ae13344065545b5f8d35fb087150a4d3303
container_end_page 401
container_issue 1
container_start_page 393
container_title International journal of hyperthermia
container_volume 38
creator Rohan, T.
Andrasina, T.
Juza, T.
Matkulcik, P.
Červinka, D.
Svobodova, I.
Novotná, V.
Bernard, V.
Valek, V.
Goldberg, S. Nahum
description To demonstrate the feasibility of irreversible electroporation (IRE) for treating biliary metal stent occlusion in an experimental liver model. IRE was performed using an expandable tubular IRE-catheter placed in nitinol stents in the porcine liver. A 3-electrode IRE-catheter was connected to an IRE-generator and one hundred 100μs pulses of constant voltage (300, 650, 1000, and 1300 V) were applied. Stent occlusion was simulated by insertion of liver tissue both ex vivo (n = 94) and in vivo in 3 pigs (n = 14). Three scenarios of the relationship between the stent, electrodes, and inserted tissue (double contact, single contact, and stent mesh contact) were studied. Electric current was measured and resistance and power calculated. Pigs were sacrificed 72 h post-procedure. Harvested samples (14 experimental, 13 controls) underwent histopathological analysis. IRE application was feasible at 300 and 650 V for the single and double contact setup in both ex vivo and in vivo studies. Significant differences in calculated resistance between double contact and single contact settings were observed (ex-vivo p ˂ 0.0001, in-vivo p = 0.02; Mann-Whitney). A mild temperature increase of the surrounding liver parenchyma was noted with increasing voltage (0.9-5.9 °C for 300-1000 V). The extent of necrotic changes in experimental samples in vivo correlated with the measured electric current (r 2 = 0.39, p = 0.01). No complications were observed during or after the in-vivo procedure. Endoluminal IRE using an expandable tubular catheter in simulated metal stent occlusion is feasible. The relationship of active catheter electrodes to stent ingrowth tissue can be estimated based on resistance values.
doi_str_mv 10.1080/02656736.2021.1875061
format article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_crossref_primary_10_1080_02656736_2021_1875061</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><doaj_id>oai_doaj_org_article_2e720148eb624c8f80aa250deab1c92c</doaj_id><sourcerecordid>2498996748</sourcerecordid><originalsourceid>FETCH-LOGICAL-c479t-c1973a6da200f0ad06e751169199f3ae13344065545b5f8d35fb087150a4d3303</originalsourceid><addsrcrecordid>eNp9kU1v1DAQhiMEokvhJ4B85JLF33FuoKpApUpc4GxN7Elx5cSL7RQKf56E3fbIydL4mffV6Gma14zuGTX0HeVa6U7oPaec7ZnpFNXsSbNjUstWMdU9bXYb027QWfOilFtKqVS8e96cCaENV4btmj-Xvw6Yw4RzhUim5DGSNJLkXFw8ejKEGCDfkwm3_1JXjmR0MEMMv6GGNJOlhPmGhJzxDnMJQ0SCEV3N6ZDyEbkLQIDUZVgiZOKgfseK-WXzbIRY8NXpPW--fbz8evG5vf7y6eriw3XrZNfX1rG-E6A9cEpHCp5q7BRjumd9PwpAJoSUVCsl1aBG44UaB2o6pihILwQV583VMdcnuLWH9dr1Ipsg2H-DlG8s5BpcRMux45RJg4Pm0pnRUACuqEcYmOu5W7PeHrMOOf1YsFQ7heIwRpgxLcVy2Zu-1500K6qOqMuplIzjYzWjdnNoHxzazaE9OVz33pwqlmFC_7j1IG0F3h-BMI8pT_Az5ehthfuY8phhdqFY8f-OvzDbrKc</addsrcrecordid><sourcetype>Open Website</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2498996748</pqid></control><display><type>article</type><title>Experimental model of occluded biliary metal stent recanalization using irreversible electroporation via a tubular catheter</title><source>Taylor &amp; Francis Open Access</source><creator>Rohan, T. ; Andrasina, T. ; Juza, T. ; Matkulcik, P. ; Červinka, D. ; Svobodova, I. ; Novotná, V. ; Bernard, V. ; Valek, V. ; Goldberg, S. Nahum</creator><creatorcontrib>Rohan, T. ; Andrasina, T. ; Juza, T. ; Matkulcik, P. ; Červinka, D. ; Svobodova, I. ; Novotná, V. ; Bernard, V. ; Valek, V. ; Goldberg, S. Nahum</creatorcontrib><description>To demonstrate the feasibility of irreversible electroporation (IRE) for treating biliary metal stent occlusion in an experimental liver model. IRE was performed using an expandable tubular IRE-catheter placed in nitinol stents in the porcine liver. A 3-electrode IRE-catheter was connected to an IRE-generator and one hundred 100μs pulses of constant voltage (300, 650, 1000, and 1300 V) were applied. Stent occlusion was simulated by insertion of liver tissue both ex vivo (n = 94) and in vivo in 3 pigs (n = 14). Three scenarios of the relationship between the stent, electrodes, and inserted tissue (double contact, single contact, and stent mesh contact) were studied. Electric current was measured and resistance and power calculated. Pigs were sacrificed 72 h post-procedure. Harvested samples (14 experimental, 13 controls) underwent histopathological analysis. IRE application was feasible at 300 and 650 V for the single and double contact setup in both ex vivo and in vivo studies. Significant differences in calculated resistance between double contact and single contact settings were observed (ex-vivo p ˂ 0.0001, in-vivo p = 0.02; Mann-Whitney). A mild temperature increase of the surrounding liver parenchyma was noted with increasing voltage (0.9-5.9 °C for 300-1000 V). The extent of necrotic changes in experimental samples in vivo correlated with the measured electric current (r 2 = 0.39, p = 0.01). No complications were observed during or after the in-vivo procedure. Endoluminal IRE using an expandable tubular catheter in simulated metal stent occlusion is feasible. The relationship of active catheter electrodes to stent ingrowth tissue can be estimated based on resistance values.</description><identifier>ISSN: 0265-6736</identifier><identifier>EISSN: 1464-5157</identifier><identifier>DOI: 10.1080/02656736.2021.1875061</identifier><identifier>PMID: 33682581</identifier><language>eng</language><publisher>England: Taylor &amp; Francis</publisher><subject>experimental model ; Irreversible electroporation ; metal stent occlusion ; metal stent recanalization ; thermocamera</subject><ispartof>International journal of hyperthermia, 2021-01, Vol.38 (1), p.393-401</ispartof><rights>2021 The Author(s). Published with license by Taylor &amp; Francis Group, LLC 2021</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c479t-c1973a6da200f0ad06e751169199f3ae13344065545b5f8d35fb087150a4d3303</citedby><cites>FETCH-LOGICAL-c479t-c1973a6da200f0ad06e751169199f3ae13344065545b5f8d35fb087150a4d3303</cites><orcidid>0000-0001-7428-1202 ; 0000-0003-0772-2287 ; 0000-0001-5723-5175 ; 0000-0003-2940-9822 ; 0000-0001-7234-8805 ; 0000-0003-0408-2771 ; 0000-0002-7105-583X ; 0000-0002-1593-0142</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://www.tandfonline.com/doi/pdf/10.1080/02656736.2021.1875061$$EPDF$$P50$$Ginformaworld$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.tandfonline.com/doi/full/10.1080/02656736.2021.1875061$$EHTML$$P50$$Ginformaworld$$Hfree_for_read</linktohtml><link.rule.ids>314,776,780,4010,27479,27900,27901,27902,59116,59117</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/33682581$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Rohan, T.</creatorcontrib><creatorcontrib>Andrasina, T.</creatorcontrib><creatorcontrib>Juza, T.</creatorcontrib><creatorcontrib>Matkulcik, P.</creatorcontrib><creatorcontrib>Červinka, D.</creatorcontrib><creatorcontrib>Svobodova, I.</creatorcontrib><creatorcontrib>Novotná, V.</creatorcontrib><creatorcontrib>Bernard, V.</creatorcontrib><creatorcontrib>Valek, V.</creatorcontrib><creatorcontrib>Goldberg, S. Nahum</creatorcontrib><title>Experimental model of occluded biliary metal stent recanalization using irreversible electroporation via a tubular catheter</title><title>International journal of hyperthermia</title><addtitle>Int J Hyperthermia</addtitle><description>To demonstrate the feasibility of irreversible electroporation (IRE) for treating biliary metal stent occlusion in an experimental liver model. IRE was performed using an expandable tubular IRE-catheter placed in nitinol stents in the porcine liver. A 3-electrode IRE-catheter was connected to an IRE-generator and one hundred 100μs pulses of constant voltage (300, 650, 1000, and 1300 V) were applied. Stent occlusion was simulated by insertion of liver tissue both ex vivo (n = 94) and in vivo in 3 pigs (n = 14). Three scenarios of the relationship between the stent, electrodes, and inserted tissue (double contact, single contact, and stent mesh contact) were studied. Electric current was measured and resistance and power calculated. Pigs were sacrificed 72 h post-procedure. Harvested samples (14 experimental, 13 controls) underwent histopathological analysis. IRE application was feasible at 300 and 650 V for the single and double contact setup in both ex vivo and in vivo studies. Significant differences in calculated resistance between double contact and single contact settings were observed (ex-vivo p ˂ 0.0001, in-vivo p = 0.02; Mann-Whitney). A mild temperature increase of the surrounding liver parenchyma was noted with increasing voltage (0.9-5.9 °C for 300-1000 V). The extent of necrotic changes in experimental samples in vivo correlated with the measured electric current (r 2 = 0.39, p = 0.01). No complications were observed during or after the in-vivo procedure. Endoluminal IRE using an expandable tubular catheter in simulated metal stent occlusion is feasible. The relationship of active catheter electrodes to stent ingrowth tissue can be estimated based on resistance values.</description><subject>experimental model</subject><subject>Irreversible electroporation</subject><subject>metal stent occlusion</subject><subject>metal stent recanalization</subject><subject>thermocamera</subject><issn>0265-6736</issn><issn>1464-5157</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><sourceid>0YH</sourceid><sourceid>DOA</sourceid><recordid>eNp9kU1v1DAQhiMEokvhJ4B85JLF33FuoKpApUpc4GxN7Elx5cSL7RQKf56E3fbIydL4mffV6Gma14zuGTX0HeVa6U7oPaec7ZnpFNXsSbNjUstWMdU9bXYb027QWfOilFtKqVS8e96cCaENV4btmj-Xvw6Yw4RzhUim5DGSNJLkXFw8ejKEGCDfkwm3_1JXjmR0MEMMv6GGNJOlhPmGhJzxDnMJQ0SCEV3N6ZDyEbkLQIDUZVgiZOKgfseK-WXzbIRY8NXpPW--fbz8evG5vf7y6eriw3XrZNfX1rG-E6A9cEpHCp5q7BRjumd9PwpAJoSUVCsl1aBG44UaB2o6pihILwQV583VMdcnuLWH9dr1Ipsg2H-DlG8s5BpcRMux45RJg4Pm0pnRUACuqEcYmOu5W7PeHrMOOf1YsFQ7heIwRpgxLcVy2Zu-1500K6qOqMuplIzjYzWjdnNoHxzazaE9OVz33pwqlmFC_7j1IG0F3h-BMI8pT_Az5ehthfuY8phhdqFY8f-OvzDbrKc</recordid><startdate>20210101</startdate><enddate>20210101</enddate><creator>Rohan, T.</creator><creator>Andrasina, T.</creator><creator>Juza, T.</creator><creator>Matkulcik, P.</creator><creator>Červinka, D.</creator><creator>Svobodova, I.</creator><creator>Novotná, V.</creator><creator>Bernard, V.</creator><creator>Valek, V.</creator><creator>Goldberg, S. Nahum</creator><general>Taylor &amp; Francis</general><general>Taylor &amp; Francis Group</general><scope>0YH</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><scope>DOA</scope><orcidid>https://orcid.org/0000-0001-7428-1202</orcidid><orcidid>https://orcid.org/0000-0003-0772-2287</orcidid><orcidid>https://orcid.org/0000-0001-5723-5175</orcidid><orcidid>https://orcid.org/0000-0003-2940-9822</orcidid><orcidid>https://orcid.org/0000-0001-7234-8805</orcidid><orcidid>https://orcid.org/0000-0003-0408-2771</orcidid><orcidid>https://orcid.org/0000-0002-7105-583X</orcidid><orcidid>https://orcid.org/0000-0002-1593-0142</orcidid></search><sort><creationdate>20210101</creationdate><title>Experimental model of occluded biliary metal stent recanalization using irreversible electroporation via a tubular catheter</title><author>Rohan, T. ; Andrasina, T. ; Juza, T. ; Matkulcik, P. ; Červinka, D. ; Svobodova, I. ; Novotná, V. ; Bernard, V. ; Valek, V. ; Goldberg, S. Nahum</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c479t-c1973a6da200f0ad06e751169199f3ae13344065545b5f8d35fb087150a4d3303</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2021</creationdate><topic>experimental model</topic><topic>Irreversible electroporation</topic><topic>metal stent occlusion</topic><topic>metal stent recanalization</topic><topic>thermocamera</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Rohan, T.</creatorcontrib><creatorcontrib>Andrasina, T.</creatorcontrib><creatorcontrib>Juza, T.</creatorcontrib><creatorcontrib>Matkulcik, P.</creatorcontrib><creatorcontrib>Červinka, D.</creatorcontrib><creatorcontrib>Svobodova, I.</creatorcontrib><creatorcontrib>Novotná, V.</creatorcontrib><creatorcontrib>Bernard, V.</creatorcontrib><creatorcontrib>Valek, V.</creatorcontrib><creatorcontrib>Goldberg, S. Nahum</creatorcontrib><collection>Taylor &amp; Francis Open Access</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><collection>Open Access: DOAJ - Directory of Open Access Journals</collection><jtitle>International journal of hyperthermia</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Rohan, T.</au><au>Andrasina, T.</au><au>Juza, T.</au><au>Matkulcik, P.</au><au>Červinka, D.</au><au>Svobodova, I.</au><au>Novotná, V.</au><au>Bernard, V.</au><au>Valek, V.</au><au>Goldberg, S. Nahum</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Experimental model of occluded biliary metal stent recanalization using irreversible electroporation via a tubular catheter</atitle><jtitle>International journal of hyperthermia</jtitle><addtitle>Int J Hyperthermia</addtitle><date>2021-01-01</date><risdate>2021</risdate><volume>38</volume><issue>1</issue><spage>393</spage><epage>401</epage><pages>393-401</pages><issn>0265-6736</issn><eissn>1464-5157</eissn><abstract>To demonstrate the feasibility of irreversible electroporation (IRE) for treating biliary metal stent occlusion in an experimental liver model. IRE was performed using an expandable tubular IRE-catheter placed in nitinol stents in the porcine liver. A 3-electrode IRE-catheter was connected to an IRE-generator and one hundred 100μs pulses of constant voltage (300, 650, 1000, and 1300 V) were applied. Stent occlusion was simulated by insertion of liver tissue both ex vivo (n = 94) and in vivo in 3 pigs (n = 14). Three scenarios of the relationship between the stent, electrodes, and inserted tissue (double contact, single contact, and stent mesh contact) were studied. Electric current was measured and resistance and power calculated. Pigs were sacrificed 72 h post-procedure. Harvested samples (14 experimental, 13 controls) underwent histopathological analysis. IRE application was feasible at 300 and 650 V for the single and double contact setup in both ex vivo and in vivo studies. Significant differences in calculated resistance between double contact and single contact settings were observed (ex-vivo p ˂ 0.0001, in-vivo p = 0.02; Mann-Whitney). A mild temperature increase of the surrounding liver parenchyma was noted with increasing voltage (0.9-5.9 °C for 300-1000 V). The extent of necrotic changes in experimental samples in vivo correlated with the measured electric current (r 2 = 0.39, p = 0.01). No complications were observed during or after the in-vivo procedure. Endoluminal IRE using an expandable tubular catheter in simulated metal stent occlusion is feasible. The relationship of active catheter electrodes to stent ingrowth tissue can be estimated based on resistance values.</abstract><cop>England</cop><pub>Taylor &amp; Francis</pub><pmid>33682581</pmid><doi>10.1080/02656736.2021.1875061</doi><tpages>9</tpages><orcidid>https://orcid.org/0000-0001-7428-1202</orcidid><orcidid>https://orcid.org/0000-0003-0772-2287</orcidid><orcidid>https://orcid.org/0000-0001-5723-5175</orcidid><orcidid>https://orcid.org/0000-0003-2940-9822</orcidid><orcidid>https://orcid.org/0000-0001-7234-8805</orcidid><orcidid>https://orcid.org/0000-0003-0408-2771</orcidid><orcidid>https://orcid.org/0000-0002-7105-583X</orcidid><orcidid>https://orcid.org/0000-0002-1593-0142</orcidid><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 0265-6736
ispartof International journal of hyperthermia, 2021-01, Vol.38 (1), p.393-401
issn 0265-6736
1464-5157
language eng
recordid cdi_crossref_primary_10_1080_02656736_2021_1875061
source Taylor & Francis Open Access
subjects experimental model
Irreversible electroporation
metal stent occlusion
metal stent recanalization
thermocamera
title Experimental model of occluded biliary metal stent recanalization using irreversible electroporation via a tubular catheter
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-08T06%3A46%3A29IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Experimental%20model%20of%20occluded%20biliary%20metal%20stent%20recanalization%20using%20irreversible%20electroporation%20via%20a%20tubular%20catheter&rft.jtitle=International%20journal%20of%20hyperthermia&rft.au=Rohan,%20T.&rft.date=2021-01-01&rft.volume=38&rft.issue=1&rft.spage=393&rft.epage=401&rft.pages=393-401&rft.issn=0265-6736&rft.eissn=1464-5157&rft_id=info:doi/10.1080/02656736.2021.1875061&rft_dat=%3Cproquest_cross%3E2498996748%3C/proquest_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c479t-c1973a6da200f0ad06e751169199f3ae13344065545b5f8d35fb087150a4d3303%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2498996748&rft_id=info:pmid/33682581&rfr_iscdi=true