Reversed Phase Liquid Chromatographic Method for the High-Throughput Analysis of Clopidogrel in Pharmaceutical Formulations Using a Monolithic Silica Column
A high-throughput, high performance liquid chromatographic method was developed and validated for the determination of clopidogrel in pharmaceutical dosage forms. The analysis was performed at room temperature using a reversed phase monolithic silica column Chromolith Performance 18e (100 mm×4.6 mm...
Saved in:
Published in: | Journal of liquid chromatography & related technologies 2005-05, Vol.28 (9), p.1357-1365 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A high-throughput, high performance liquid chromatographic method was developed and validated for the determination of clopidogrel in pharmaceutical dosage forms. The analysis was performed at room temperature using a reversed phase monolithic silica column Chromolith Performance 18e (100 mm×4.6 mm I.D.). The mobile phase consisted of acetonitrile:phosphate buffer (50∶50 v/v, pH 3.0) at a flow rate of 4.0 mL/min. The photodiode array detector was set at 235 nm. The developed method showed a good linear relationship in the concentration range from 1.0 to 40.0 µg/mL with a correlation coefficient of 0.999. The limit of detection and limit of quantification were 0.97 µg/mL and 3.52 µg/mL, respectively. |
---|---|
ISSN: | 1082-6076 1520-572X |