Comparison of air kerma measurements for tungsten anode based mammography x-ray beam qualities (EURAMET.RI(I)-S4.1)
A comparison of the air kerma standards for x-radiation qualities used in mammography was performed between the PTB and the IAEA. Two reference-class ionization chamber types Radcal RC6M and Magna A650 of the IAEA and tungsten anode based beam qualities with Mo and Al external filtrations ( W+Mo , W...
Saved in:
Published in: | Metrologia 2016-01, Vol.53 (1A), p.6005 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A comparison of the air kerma standards for x-radiation qualities used in mammography was performed between the PTB and the IAEA. Two reference-class ionization chamber types Radcal RC6M and Magna A650 of the IAEA and tungsten anode based beam qualities with Mo and Al external filtrations (
W+Mo
,
W+Al
) established at both laboratories were selected for the comparison. The calibration coefficients,
N
K
_air
, were determined for the transfer chambers at the PTB in May 2015 and before and after this at the IAEA Dosimetry Laboratory. The results show good agreement, to be well within the 0.55 % standard uncertainty of the comparison. Correction factors to determine
N
K
_air
for these beam qualities based on calibration in RQR-M mammography beam qualities, established according to the IEC 61267 standard, were also calculated for the Radcal RC6M, 10X5-6M, and Magna A650 types of chambers.
Main text
To reach the main text of this paper, click on
Final Report
. Note that this text is that which appears in Appendix B of the BIPM key comparison database
kcdb.bipm.org/
.
The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA). |
---|---|
ISSN: | 0026-1394 1681-7575 |
DOI: | 10.1088/0026-1394/53/1A/06005 |