Diamond is widely adopted in physics experiments owing to its good radiation hardness. In this paper, a low noise electronics system based on charge sensitive amplifier (CSA) is designed to read out diamond microstrip detectors. Up to 40 channels are implemented in the system, each containing a CSA,...

Full description

Saved in:
Bibliographic Details
Published in:Journal of physics. Conference series 2022-11, Vol.2374 (1), p.12079
Main Authors: Luo, Laifu, Shen, Zhongtao, Yu, Hanlin, Zhang, Jianyong, Liu, Shubin
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Diamond is widely adopted in physics experiments owing to its good radiation hardness. In this paper, a low noise electronics system based on charge sensitive amplifier (CSA) is designed to read out diamond microstrip detectors. Up to 40 channels are implemented in the system, each containing a CSA, a CR — RC 2 shaper, an analog to digital convertor (ADC) and a discriminator used for trigger generation. After calibration and joint test with prototype detector with a size of 4×4×0.5 mm 3 , a noise level of less than 845 electrons is realized in all 40 channels. Furthermore, the result with 90 Sr radiation source is consistent with the theory, and a position resolution of 13 um is obtained in laser test which indicates the readout system meets the demand of the future experiment.