Loading…
Accuracy Advancement of Non-Contact Temperature Measurement with Differential Technology
Generally, accuracy of non-contact temperature measurement is not high because of the effects of ambient temperature and surface thermal emissivity of measured target. Differential measurement technology is measuring both temperature of a reference target and a measured target with the same temperat...
Saved in:
Published in: | Journal of physics. Conference series 2006-10, Vol.48 (1), p.720-724 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Generally, accuracy of non-contact temperature measurement is not high because of the effects of ambient temperature and surface thermal emissivity of measured target. Differential measurement technology is measuring both temperature of a reference target and a measured target with the same temperature measurement device. As ambient temperature, rays and surface thermal emissivity of heating unit have nearly the same effects on two targets, actual measurement error of reference temperature actual value can be used to correct temperature error of measured target. When working environment of reference target and measured target are basically the same, differential measurement can eliminate most measurement error caused by environmental factors. This method is mainly used in non-contact temperature measurement requiring high accuracy. |
---|---|
ISSN: | 1742-6596 1742-6588 1742-6596 |
DOI: | 10.1088/1742-6596/48/1/136 |