P2855Unexpectedly high rate of lead failure of the Microport (formerly Sorin/Livanova) Beflex and Vega pacemaker electrodes: A single centre experience

Abstract Introduction Pacing leads remain the weak link of current pacemaker systems. Various differences in design and material exist among companies. Lead performance is mainly assessed via post-marketing studies of the manufacturing companies. Reliable independent reports are rare. We aimed to st...

Full description

Saved in:
Bibliographic Details
Published in:European heart journal 2019-10, Vol.40 (Supplement_1)
Main Authors: Anwander, M, Kueffer, T, Branca, M, Sweda, R, Nozica, N, Asatryan, B, Lam, A, Baldinger, S, Haeberlin, A, Seiler, J, Servatius, H, Tanner, H, Reichlin, T, Roten, L, Noti, F
Format: Article
Language:English
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c1394-e886506e3198be06f3bae41ff94dff6cb92f8f8442547faddb5d5118dfe71b23
cites
container_end_page
container_issue Supplement_1
container_start_page
container_title European heart journal
container_volume 40
creator Anwander, M
Kueffer, T
Branca, M
Sweda, R
Nozica, N
Asatryan, B
Lam, A
Baldinger, S
Haeberlin, A
Seiler, J
Servatius, H
Tanner, H
Reichlin, T
Roten, L
Noti, F
description Abstract Introduction Pacing leads remain the weak link of current pacemaker systems. Various differences in design and material exist among companies. Lead performance is mainly assessed via post-marketing studies of the manufacturing companies. Reliable independent reports are rare. We aimed to study the early and long-term performance of the Microport (formerly Sorin/Livanova) Beflex and Vega leads at our centre, for which a lead survival >99% at 3 years has been reported by the company. Method In this single centre, retrospective study we analysed the performance of all right ventricular Microport pacemaker leads implanted at our centre between January 2014 and January 2018. Only first pacemaker implants were considered. Lead failure was defined as any lead issue requiring reintervention during follow-up (dislocation, perforation, electrical abnormalities such as lead noise or excessively high thresholds). Results A total of 271 Microport right ventricular pacing leads were implanted (233 Beflex and 38 Vega leads). Mean patient age was 76±13.1 years (66% men). Dual chamber pacemakers were implanted in 162 patients (60%) and single chamber in 109 (40%). Mean threshold at implant was 0.6V/0.5ms (range 0.3–1.2V), mean R wave 13.2 mV (range 1.5–30mV) and mean impedance 816 Ohm (range 469–1639 Ohm). Patients without available follow-up information were excluded (N=18, 6.6%). The remaining 253 patients (93.4%) were analysed. Median follow-up was 1.26 years, IQR [25%=0.91 and 75%=2.24]. We observed a total of 25 lead failures (10%). Lead dislocation occurred in 2 cases (0.8%), lead perforation in 5 cases (2%), electrical abnormalities in 6 cases (2.4%) and excessively high threshold in 12 cases (4.8%; mean voltage 4V, range 2–7.5V; mean pulse width 0.75ms, range 0.35–1ms). Yearly incidence of lead failure per 100 leads was 6.1% (95%-CI [4.09–8.98] with a failure rate of 12.74% at 3 year in Kaplan-Meier analysis (Figure). Figure 1 Conclusion We found an unexpectedly high rate of lead failure of the Microport Beflex and Vega pacing leads at our centre. The two main reasons for premature lead failure were excessively high thresholds as well as electrical abnormalities during follow-up. Comparison of lead performance with other centres and against other leads are needed to further assess the magnitude of the problem.
doi_str_mv 10.1093/eurheartj/ehz748.1164
format article
fullrecord <record><control><sourceid>oup_cross</sourceid><recordid>TN_cdi_crossref_primary_10_1093_eurheartj_ehz748_1164</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><oup_id>10.1093/eurheartj/ehz748.1164</oup_id><sourcerecordid>10.1093/eurheartj/ehz748.1164</sourcerecordid><originalsourceid>FETCH-LOGICAL-c1394-e886506e3198be06f3bae41ff94dff6cb92f8f8442547faddb5d5118dfe71b23</originalsourceid><addsrcrecordid>eNqNkMtOwzAQRS0EEqXwCUhewiKtndiuza5UvKQikCiIXeTE48YljSMnrVp-hN8lpYg1q9Eszr0zB6FzSgaUqGQIq1CADu1iCMXniMkBpYIdoB7lcRwpwfgh6hGqeCSEfD9GJ02zIIRIQUUPfT3HkvPXCjY15C2YcosLNy9w0C1gb3EJ2mCrXbkKP3tbAH50efC1Dy2-sD4sIXTQiw-uGk7dWld-rS_xNdgSNlhXBr_BXONa57DUHxAwlF1R8AaaKzzGjavmJeAcqrYr2F0RHFQ5nKIjq8sGzn5nH81ub2aT-2j6dPcwGU-jnCaKRSCl4ERAQpXMgAibZBoYtVYxY63IMxVbaSVjMWcjq43JuOGUSmNhRLM46SO-j-0-apoANq2DW-qwTSlJd3LTP7npXm66k9txZM_5Vf1P5Bv-Z4UH</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype></control><display><type>article</type><title>P2855Unexpectedly high rate of lead failure of the Microport (formerly Sorin/Livanova) Beflex and Vega pacemaker electrodes: A single centre experience</title><source>Oxford Journals Online</source><creator>Anwander, M ; Kueffer, T ; Branca, M ; Sweda, R ; Nozica, N ; Asatryan, B ; Lam, A ; Baldinger, S ; Haeberlin, A ; Seiler, J ; Servatius, H ; Tanner, H ; Reichlin, T ; Roten, L ; Noti, F</creator><creatorcontrib>Anwander, M ; Kueffer, T ; Branca, M ; Sweda, R ; Nozica, N ; Asatryan, B ; Lam, A ; Baldinger, S ; Haeberlin, A ; Seiler, J ; Servatius, H ; Tanner, H ; Reichlin, T ; Roten, L ; Noti, F</creatorcontrib><description>Abstract Introduction Pacing leads remain the weak link of current pacemaker systems. Various differences in design and material exist among companies. Lead performance is mainly assessed via post-marketing studies of the manufacturing companies. Reliable independent reports are rare. We aimed to study the early and long-term performance of the Microport (formerly Sorin/Livanova) Beflex and Vega leads at our centre, for which a lead survival &gt;99% at 3 years has been reported by the company. Method In this single centre, retrospective study we analysed the performance of all right ventricular Microport pacemaker leads implanted at our centre between January 2014 and January 2018. Only first pacemaker implants were considered. Lead failure was defined as any lead issue requiring reintervention during follow-up (dislocation, perforation, electrical abnormalities such as lead noise or excessively high thresholds). Results A total of 271 Microport right ventricular pacing leads were implanted (233 Beflex and 38 Vega leads). Mean patient age was 76±13.1 years (66% men). Dual chamber pacemakers were implanted in 162 patients (60%) and single chamber in 109 (40%). Mean threshold at implant was 0.6V/0.5ms (range 0.3–1.2V), mean R wave 13.2 mV (range 1.5–30mV) and mean impedance 816 Ohm (range 469–1639 Ohm). Patients without available follow-up information were excluded (N=18, 6.6%). The remaining 253 patients (93.4%) were analysed. Median follow-up was 1.26 years, IQR [25%=0.91 and 75%=2.24]. We observed a total of 25 lead failures (10%). Lead dislocation occurred in 2 cases (0.8%), lead perforation in 5 cases (2%), electrical abnormalities in 6 cases (2.4%) and excessively high threshold in 12 cases (4.8%; mean voltage 4V, range 2–7.5V; mean pulse width 0.75ms, range 0.35–1ms). Yearly incidence of lead failure per 100 leads was 6.1% (95%-CI [4.09–8.98] with a failure rate of 12.74% at 3 year in Kaplan-Meier analysis (Figure). Figure 1 Conclusion We found an unexpectedly high rate of lead failure of the Microport Beflex and Vega pacing leads at our centre. The two main reasons for premature lead failure were excessively high thresholds as well as electrical abnormalities during follow-up. Comparison of lead performance with other centres and against other leads are needed to further assess the magnitude of the problem.</description><identifier>ISSN: 0195-668X</identifier><identifier>EISSN: 1522-9645</identifier><identifier>DOI: 10.1093/eurheartj/ehz748.1164</identifier><language>eng</language><publisher>Oxford University Press</publisher><ispartof>European heart journal, 2019-10, Vol.40 (Supplement_1)</ispartof><rights>Published on behalf of the European Society of Cardiology. All rights reserved. © The Author(s) 2019. For permissions, please email: journals.permissions@oup.com. 2019</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c1394-e886506e3198be06f3bae41ff94dff6cb92f8f8442547faddb5d5118dfe71b23</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925</link.rule.ids></links><search><creatorcontrib>Anwander, M</creatorcontrib><creatorcontrib>Kueffer, T</creatorcontrib><creatorcontrib>Branca, M</creatorcontrib><creatorcontrib>Sweda, R</creatorcontrib><creatorcontrib>Nozica, N</creatorcontrib><creatorcontrib>Asatryan, B</creatorcontrib><creatorcontrib>Lam, A</creatorcontrib><creatorcontrib>Baldinger, S</creatorcontrib><creatorcontrib>Haeberlin, A</creatorcontrib><creatorcontrib>Seiler, J</creatorcontrib><creatorcontrib>Servatius, H</creatorcontrib><creatorcontrib>Tanner, H</creatorcontrib><creatorcontrib>Reichlin, T</creatorcontrib><creatorcontrib>Roten, L</creatorcontrib><creatorcontrib>Noti, F</creatorcontrib><title>P2855Unexpectedly high rate of lead failure of the Microport (formerly Sorin/Livanova) Beflex and Vega pacemaker electrodes: A single centre experience</title><title>European heart journal</title><description>Abstract Introduction Pacing leads remain the weak link of current pacemaker systems. Various differences in design and material exist among companies. Lead performance is mainly assessed via post-marketing studies of the manufacturing companies. Reliable independent reports are rare. We aimed to study the early and long-term performance of the Microport (formerly Sorin/Livanova) Beflex and Vega leads at our centre, for which a lead survival &gt;99% at 3 years has been reported by the company. Method In this single centre, retrospective study we analysed the performance of all right ventricular Microport pacemaker leads implanted at our centre between January 2014 and January 2018. Only first pacemaker implants were considered. Lead failure was defined as any lead issue requiring reintervention during follow-up (dislocation, perforation, electrical abnormalities such as lead noise or excessively high thresholds). Results A total of 271 Microport right ventricular pacing leads were implanted (233 Beflex and 38 Vega leads). Mean patient age was 76±13.1 years (66% men). Dual chamber pacemakers were implanted in 162 patients (60%) and single chamber in 109 (40%). Mean threshold at implant was 0.6V/0.5ms (range 0.3–1.2V), mean R wave 13.2 mV (range 1.5–30mV) and mean impedance 816 Ohm (range 469–1639 Ohm). Patients without available follow-up information were excluded (N=18, 6.6%). The remaining 253 patients (93.4%) were analysed. Median follow-up was 1.26 years, IQR [25%=0.91 and 75%=2.24]. We observed a total of 25 lead failures (10%). Lead dislocation occurred in 2 cases (0.8%), lead perforation in 5 cases (2%), electrical abnormalities in 6 cases (2.4%) and excessively high threshold in 12 cases (4.8%; mean voltage 4V, range 2–7.5V; mean pulse width 0.75ms, range 0.35–1ms). Yearly incidence of lead failure per 100 leads was 6.1% (95%-CI [4.09–8.98] with a failure rate of 12.74% at 3 year in Kaplan-Meier analysis (Figure). Figure 1 Conclusion We found an unexpectedly high rate of lead failure of the Microport Beflex and Vega pacing leads at our centre. The two main reasons for premature lead failure were excessively high thresholds as well as electrical abnormalities during follow-up. Comparison of lead performance with other centres and against other leads are needed to further assess the magnitude of the problem.</description><issn>0195-668X</issn><issn>1522-9645</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2019</creationdate><recordtype>article</recordtype><recordid>eNqNkMtOwzAQRS0EEqXwCUhewiKtndiuza5UvKQikCiIXeTE48YljSMnrVp-hN8lpYg1q9Eszr0zB6FzSgaUqGQIq1CADu1iCMXniMkBpYIdoB7lcRwpwfgh6hGqeCSEfD9GJ02zIIRIQUUPfT3HkvPXCjY15C2YcosLNy9w0C1gb3EJ2mCrXbkKP3tbAH50efC1Dy2-sD4sIXTQiw-uGk7dWld-rS_xNdgSNlhXBr_BXONa57DUHxAwlF1R8AaaKzzGjavmJeAcqrYr2F0RHFQ5nKIjq8sGzn5nH81ub2aT-2j6dPcwGU-jnCaKRSCl4ERAQpXMgAibZBoYtVYxY63IMxVbaSVjMWcjq43JuOGUSmNhRLM46SO-j-0-apoANq2DW-qwTSlJd3LTP7npXm66k9txZM_5Vf1P5Bv-Z4UH</recordid><startdate>20191001</startdate><enddate>20191001</enddate><creator>Anwander, M</creator><creator>Kueffer, T</creator><creator>Branca, M</creator><creator>Sweda, R</creator><creator>Nozica, N</creator><creator>Asatryan, B</creator><creator>Lam, A</creator><creator>Baldinger, S</creator><creator>Haeberlin, A</creator><creator>Seiler, J</creator><creator>Servatius, H</creator><creator>Tanner, H</creator><creator>Reichlin, T</creator><creator>Roten, L</creator><creator>Noti, F</creator><general>Oxford University Press</general><scope>AAYXX</scope><scope>CITATION</scope></search><sort><creationdate>20191001</creationdate><title>P2855Unexpectedly high rate of lead failure of the Microport (formerly Sorin/Livanova) Beflex and Vega pacemaker electrodes: A single centre experience</title><author>Anwander, M ; Kueffer, T ; Branca, M ; Sweda, R ; Nozica, N ; Asatryan, B ; Lam, A ; Baldinger, S ; Haeberlin, A ; Seiler, J ; Servatius, H ; Tanner, H ; Reichlin, T ; Roten, L ; Noti, F</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c1394-e886506e3198be06f3bae41ff94dff6cb92f8f8442547faddb5d5118dfe71b23</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2019</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Anwander, M</creatorcontrib><creatorcontrib>Kueffer, T</creatorcontrib><creatorcontrib>Branca, M</creatorcontrib><creatorcontrib>Sweda, R</creatorcontrib><creatorcontrib>Nozica, N</creatorcontrib><creatorcontrib>Asatryan, B</creatorcontrib><creatorcontrib>Lam, A</creatorcontrib><creatorcontrib>Baldinger, S</creatorcontrib><creatorcontrib>Haeberlin, A</creatorcontrib><creatorcontrib>Seiler, J</creatorcontrib><creatorcontrib>Servatius, H</creatorcontrib><creatorcontrib>Tanner, H</creatorcontrib><creatorcontrib>Reichlin, T</creatorcontrib><creatorcontrib>Roten, L</creatorcontrib><creatorcontrib>Noti, F</creatorcontrib><collection>CrossRef</collection><jtitle>European heart journal</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Anwander, M</au><au>Kueffer, T</au><au>Branca, M</au><au>Sweda, R</au><au>Nozica, N</au><au>Asatryan, B</au><au>Lam, A</au><au>Baldinger, S</au><au>Haeberlin, A</au><au>Seiler, J</au><au>Servatius, H</au><au>Tanner, H</au><au>Reichlin, T</au><au>Roten, L</au><au>Noti, F</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>P2855Unexpectedly high rate of lead failure of the Microport (formerly Sorin/Livanova) Beflex and Vega pacemaker electrodes: A single centre experience</atitle><jtitle>European heart journal</jtitle><date>2019-10-01</date><risdate>2019</risdate><volume>40</volume><issue>Supplement_1</issue><issn>0195-668X</issn><eissn>1522-9645</eissn><abstract>Abstract Introduction Pacing leads remain the weak link of current pacemaker systems. Various differences in design and material exist among companies. Lead performance is mainly assessed via post-marketing studies of the manufacturing companies. Reliable independent reports are rare. We aimed to study the early and long-term performance of the Microport (formerly Sorin/Livanova) Beflex and Vega leads at our centre, for which a lead survival &gt;99% at 3 years has been reported by the company. Method In this single centre, retrospective study we analysed the performance of all right ventricular Microport pacemaker leads implanted at our centre between January 2014 and January 2018. Only first pacemaker implants were considered. Lead failure was defined as any lead issue requiring reintervention during follow-up (dislocation, perforation, electrical abnormalities such as lead noise or excessively high thresholds). Results A total of 271 Microport right ventricular pacing leads were implanted (233 Beflex and 38 Vega leads). Mean patient age was 76±13.1 years (66% men). Dual chamber pacemakers were implanted in 162 patients (60%) and single chamber in 109 (40%). Mean threshold at implant was 0.6V/0.5ms (range 0.3–1.2V), mean R wave 13.2 mV (range 1.5–30mV) and mean impedance 816 Ohm (range 469–1639 Ohm). Patients without available follow-up information were excluded (N=18, 6.6%). The remaining 253 patients (93.4%) were analysed. Median follow-up was 1.26 years, IQR [25%=0.91 and 75%=2.24]. We observed a total of 25 lead failures (10%). Lead dislocation occurred in 2 cases (0.8%), lead perforation in 5 cases (2%), electrical abnormalities in 6 cases (2.4%) and excessively high threshold in 12 cases (4.8%; mean voltage 4V, range 2–7.5V; mean pulse width 0.75ms, range 0.35–1ms). Yearly incidence of lead failure per 100 leads was 6.1% (95%-CI [4.09–8.98] with a failure rate of 12.74% at 3 year in Kaplan-Meier analysis (Figure). Figure 1 Conclusion We found an unexpectedly high rate of lead failure of the Microport Beflex and Vega pacing leads at our centre. The two main reasons for premature lead failure were excessively high thresholds as well as electrical abnormalities during follow-up. Comparison of lead performance with other centres and against other leads are needed to further assess the magnitude of the problem.</abstract><pub>Oxford University Press</pub><doi>10.1093/eurheartj/ehz748.1164</doi></addata></record>
fulltext fulltext
identifier ISSN: 0195-668X
ispartof European heart journal, 2019-10, Vol.40 (Supplement_1)
issn 0195-668X
1522-9645
language eng
recordid cdi_crossref_primary_10_1093_eurheartj_ehz748_1164
source Oxford Journals Online
title P2855Unexpectedly high rate of lead failure of the Microport (formerly Sorin/Livanova) Beflex and Vega pacemaker electrodes: A single centre experience
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-28T07%3A06%3A11IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-oup_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=P2855Unexpectedly%20high%20rate%20of%20lead%20failure%20of%20the%20Microport%20(formerly%20Sorin/Livanova)%20Beflex%20and%20Vega%20pacemaker%20electrodes:%20A%20single%20centre%20experience&rft.jtitle=European%20heart%20journal&rft.au=Anwander,%20M&rft.date=2019-10-01&rft.volume=40&rft.issue=Supplement_1&rft.issn=0195-668X&rft.eissn=1522-9645&rft_id=info:doi/10.1093/eurheartj/ehz748.1164&rft_dat=%3Coup_cross%3E10.1093/eurheartj/ehz748.1164%3C/oup_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c1394-e886506e3198be06f3bae41ff94dff6cb92f8f8442547faddb5d5118dfe71b23%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_id=info:pmid/&rft_oup_id=10.1093/eurheartj/ehz748.1164&rfr_iscdi=true