Comparison of Two Methods of Obtaining the Computed Tomography Dose Index by Using Different TL Materials: LiF:Mg,Cu,P and LiF:Mg,Ti
TLD results from two methods of calculating CTDI, from a single slice dose profile and from multiple scan average dose, were compared for different slice thicknesses. Two types of TL materials, LiF:Mg,Cu,P and LiF:Mg,Ti, were employed in order to analyse the differences in energy response between th...
Saved in:
Published in: | Radiation protection dosimetry 1999-09, Vol.85 (1-4), p.429-432, Article 429 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | TLD results from two methods of calculating CTDI, from a single slice dose profile and from multiple scan average dose, were compared for different slice thicknesses. Two types of TL materials, LiF:Mg,Cu,P and LiF:Mg,Ti, were employed in order to analyse the differences in energy response between the two phosphors. Results revealed the multiple scan average dose as the more practical and realistic method and confirmed the importance of performing appropriate energy response corrections. |
---|---|