Backscatter and Attenuation of Photon Radiation by a User and Their Influence on Portable Radiation Monitor Response

The influence on the response of portable photon radiation monitors of backscatter and attenuation by the instrument user is investigated using tissue equivalent phantoms. The conclusion is that instruments whose characteristics are acceptable when measured in free air will not be unduly affected by...

Full description

Saved in:
Bibliographic Details
Published in:Radiation protection dosimetry 1984-12, Vol.9 (4), p.263-268
Main Author: BURGESS, P. H
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The influence on the response of portable photon radiation monitors of backscatter and attenuation by the instrument user is investigated using tissue equivalent phantoms. The conclusion is that instruments whose characteristics are acceptable when measured in free air will not be unduly affected by backscatter from the operator. There is, however, a significant attenuation of radiation incident from behind the operator.