Loading…
Two-dimensional thermoluminescence dosimetry using planar detectors and a TL reader with CCD camera readout
A novel method of determining two-dimensional (2-D) dose distributions is presented, using in-house developed, large-area (a few cm2) thermoluminescent (TL) detectors based on LiF powder plated on Al foil. An in-house developed planar large-area TL reader equipped with a coupled charge device (CCD)...
Saved in:
Published in: | Radiation protection dosimetry 2006-09, Vol.120 (1-4), p.129-132 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A novel method of determining two-dimensional (2-D) dose distributions is presented, using in-house developed, large-area (a few cm2) thermoluminescent (TL) detectors based on LiF powder plated on Al foil. An in-house developed planar large-area TL reader equipped with a coupled charge device (CCD) camera is used for readout, providing digital images of 2-D dose distributions on the surface of these large-area TL detectors. The capability of the newly developed system is demonstrated by mapping 2-D dose distributions around a brachytherapy source, at dose ranges and source geometries relevant for clinical radiotherapy. Examples of local and dynamic evaluation of TL output from conventional TL detectors are also shown. |
---|---|
ISSN: | 0144-8420 1742-3406 |
DOI: | 10.1093/rpd/nci593 |