Measurement of oxygen uptake and carbon dioxide elimination using the bymixer: Validation in a metabolic lung simulator

The authors have developed a new clinical bymixer that bypasses a constant fraction of gas flow through a mixing arm. A separate bymixer was interposed in the expiratory and inspiratory limbs of the ventilation circuit to measure mixed gas fractions. By utilizing nitrogen conservation, the clinical...

Full description

Saved in:
Bibliographic Details
Published in:Anesthesiology (Philadelphia) 2004-06, Vol.100 (6), p.1427-1437
Main Authors: ROSENBAUM, Abraham, KIRBY, Christopher, BREEN, Peter H
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by
cites
container_end_page 1437
container_issue 6
container_start_page 1427
container_title Anesthesiology (Philadelphia)
container_volume 100
creator ROSENBAUM, Abraham
KIRBY, Christopher
BREEN, Peter H
description The authors have developed a new clinical bymixer that bypasses a constant fraction of gas flow through a mixing arm. A separate bymixer was interposed in the expiratory and inspiratory limbs of the ventilation circuit to measure mixed gas fractions. By utilizing nitrogen conservation, the clinical bymixer allows the determination of airway carbon dioxide elimination (VCO2) and oxygen uptake (VO2), whenever basic expired flow and gas monitoring measurements are used for the patient. Neither an expiratory exhaust gas collection bag nor expensive, complex equipment are needed. This study tested the accuracy of airway bymixer-flow measurements of VCO2 and VO2 in a new bench apparatus. The authors compared airway bymixer-flow measurements of VCO2 and VO2 over a range of reference values generated by ethanol combustion in a new metabolic lung simulator, which was ventilated by a volume-cycled respirator. An airway humidity and temperature sensor permitted standard temperature and pressure, dry, correction of airway VCO2 and VO2. Bymixer-flow airway measurements of VCO2 and VO2 correlated closely (R2 = 0.999 and 0.998, respectively) with the stoichiometric values generated by ethanol combustion. Limits of agreement for VCO2 and VO2 were 0.1 +/- 4.7 and 1.1 +/- 5.7%, respectively. The average (+/- SD) percent error for airway VCO2 (compared with the stoichiometric value) was 0.1 +/- 2.4%. The same error for airway VO2 was 1.1 +/- 2.9%. The new clinical bymixer, plus basic expired flow and gas fraction measurements, generated clinically accurate determinations of VCO2 and VO2. These measurements are helpful in the assessment of metabolic gas exchange in the critical care unit. In contrast to using the gas collection bag or complex metabolic monitor, the bymixer should measure mixed gas concentrations in the inspired or expired limb of the common anesthesia circle ventilation circuit.
doi_str_mv 10.1097/00000542-200406000-00015
format article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_crossref_primary_10_1097_00000542_200406000_00015</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>71956907</sourcerecordid><originalsourceid>FETCH-LOGICAL-c286t-2405022a09e628cdcb8a93773d0e195fc9958011dc90ecf84dae7692159f38ba3</originalsourceid><addsrcrecordid>eNpNkEuPFCEQgInRuLOrf8Fw0Vsr0MPLm9msuskaL-q1Uw3VK0rDCN1x5t_LOOODhJAqvqqCjxDK2UvOrH7FjktuRScY2zLVgq5tLh-QDZfCdJxr-ZBsWq7veibEBbms9VsLtezNY3LBJVdKKrEhPz8g1LXgjGmheaJ5f7jHRNfdAt-RQvLUQRlzoj7kffBIMYY5JFhCy601pHu6fEU6Huawx_KafoEY_Ok2JAp0xgXGHIOjcW1sDfMaYcnlCXk0Qaz49Hxekc9vbz5dv-_uPr67vX5z1zlh1NKJLZPt_cAsKmGcd6MB22vde4bcyslZKw3j3DvL0E1m6wG1soJLO_VmhP6KvDj13ZX8Y8W6DHOoDmOEhHmtg25dlGW6geYEupJrLTgNuxJmKIeBs-EoffgjffgrffgtvZU-O89Yxxn9v8Kz5QY8PwNQHcSpQHKh_sfp459U_wtAiosb</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>71956907</pqid></control><display><type>article</type><title>Measurement of oxygen uptake and carbon dioxide elimination using the bymixer: Validation in a metabolic lung simulator</title><source>HEAL-Link subscriptions: Lippincott Williams &amp; Wilkins</source><creator>ROSENBAUM, Abraham ; KIRBY, Christopher ; BREEN, Peter H</creator><creatorcontrib>ROSENBAUM, Abraham ; KIRBY, Christopher ; BREEN, Peter H</creatorcontrib><description>The authors have developed a new clinical bymixer that bypasses a constant fraction of gas flow through a mixing arm. A separate bymixer was interposed in the expiratory and inspiratory limbs of the ventilation circuit to measure mixed gas fractions. By utilizing nitrogen conservation, the clinical bymixer allows the determination of airway carbon dioxide elimination (VCO2) and oxygen uptake (VO2), whenever basic expired flow and gas monitoring measurements are used for the patient. Neither an expiratory exhaust gas collection bag nor expensive, complex equipment are needed. This study tested the accuracy of airway bymixer-flow measurements of VCO2 and VO2 in a new bench apparatus. The authors compared airway bymixer-flow measurements of VCO2 and VO2 over a range of reference values generated by ethanol combustion in a new metabolic lung simulator, which was ventilated by a volume-cycled respirator. An airway humidity and temperature sensor permitted standard temperature and pressure, dry, correction of airway VCO2 and VO2. Bymixer-flow airway measurements of VCO2 and VO2 correlated closely (R2 = 0.999 and 0.998, respectively) with the stoichiometric values generated by ethanol combustion. Limits of agreement for VCO2 and VO2 were 0.1 +/- 4.7 and 1.1 +/- 5.7%, respectively. The average (+/- SD) percent error for airway VCO2 (compared with the stoichiometric value) was 0.1 +/- 2.4%. The same error for airway VO2 was 1.1 +/- 2.9%. The new clinical bymixer, plus basic expired flow and gas fraction measurements, generated clinically accurate determinations of VCO2 and VO2. These measurements are helpful in the assessment of metabolic gas exchange in the critical care unit. In contrast to using the gas collection bag or complex metabolic monitor, the bymixer should measure mixed gas concentrations in the inspired or expired limb of the common anesthesia circle ventilation circuit.</description><identifier>ISSN: 0003-3022</identifier><identifier>EISSN: 1528-1175</identifier><identifier>DOI: 10.1097/00000542-200406000-00015</identifier><identifier>PMID: 15166562</identifier><identifier>CODEN: ANESAV</identifier><language>eng</language><publisher>Hagerstown, MD: Lippincott</publisher><subject>Anesthesia ; Anesthesia, Closed-Circuit - instrumentation ; Anesthesia, Closed-Circuit - methods ; Anesthesia. Intensive care medicine. Transfusions. Cell therapy and gene therapy ; Biological and medical sciences ; Breath Tests - instrumentation ; Breath Tests - methods ; Carbon Dioxide - analysis ; Carbon Dioxide - metabolism ; Humans ; In Vitro Techniques ; Lung - metabolism ; Medical sciences ; Oxygen Consumption - physiology ; Pulmonary Gas Exchange ; Respiration, Artificial - instrumentation ; Respiration, Artificial - methods ; Ventilators, Mechanical - standards</subject><ispartof>Anesthesiology (Philadelphia), 2004-06, Vol.100 (6), p.1427-1437</ispartof><rights>2004 INIST-CNRS</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,27903,27904</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&amp;idt=15793776$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/15166562$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>ROSENBAUM, Abraham</creatorcontrib><creatorcontrib>KIRBY, Christopher</creatorcontrib><creatorcontrib>BREEN, Peter H</creatorcontrib><title>Measurement of oxygen uptake and carbon dioxide elimination using the bymixer: Validation in a metabolic lung simulator</title><title>Anesthesiology (Philadelphia)</title><addtitle>Anesthesiology</addtitle><description>The authors have developed a new clinical bymixer that bypasses a constant fraction of gas flow through a mixing arm. A separate bymixer was interposed in the expiratory and inspiratory limbs of the ventilation circuit to measure mixed gas fractions. By utilizing nitrogen conservation, the clinical bymixer allows the determination of airway carbon dioxide elimination (VCO2) and oxygen uptake (VO2), whenever basic expired flow and gas monitoring measurements are used for the patient. Neither an expiratory exhaust gas collection bag nor expensive, complex equipment are needed. This study tested the accuracy of airway bymixer-flow measurements of VCO2 and VO2 in a new bench apparatus. The authors compared airway bymixer-flow measurements of VCO2 and VO2 over a range of reference values generated by ethanol combustion in a new metabolic lung simulator, which was ventilated by a volume-cycled respirator. An airway humidity and temperature sensor permitted standard temperature and pressure, dry, correction of airway VCO2 and VO2. Bymixer-flow airway measurements of VCO2 and VO2 correlated closely (R2 = 0.999 and 0.998, respectively) with the stoichiometric values generated by ethanol combustion. Limits of agreement for VCO2 and VO2 were 0.1 +/- 4.7 and 1.1 +/- 5.7%, respectively. The average (+/- SD) percent error for airway VCO2 (compared with the stoichiometric value) was 0.1 +/- 2.4%. The same error for airway VO2 was 1.1 +/- 2.9%. The new clinical bymixer, plus basic expired flow and gas fraction measurements, generated clinically accurate determinations of VCO2 and VO2. These measurements are helpful in the assessment of metabolic gas exchange in the critical care unit. In contrast to using the gas collection bag or complex metabolic monitor, the bymixer should measure mixed gas concentrations in the inspired or expired limb of the common anesthesia circle ventilation circuit.</description><subject>Anesthesia</subject><subject>Anesthesia, Closed-Circuit - instrumentation</subject><subject>Anesthesia, Closed-Circuit - methods</subject><subject>Anesthesia. Intensive care medicine. Transfusions. Cell therapy and gene therapy</subject><subject>Biological and medical sciences</subject><subject>Breath Tests - instrumentation</subject><subject>Breath Tests - methods</subject><subject>Carbon Dioxide - analysis</subject><subject>Carbon Dioxide - metabolism</subject><subject>Humans</subject><subject>In Vitro Techniques</subject><subject>Lung - metabolism</subject><subject>Medical sciences</subject><subject>Oxygen Consumption - physiology</subject><subject>Pulmonary Gas Exchange</subject><subject>Respiration, Artificial - instrumentation</subject><subject>Respiration, Artificial - methods</subject><subject>Ventilators, Mechanical - standards</subject><issn>0003-3022</issn><issn>1528-1175</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2004</creationdate><recordtype>article</recordtype><recordid>eNpNkEuPFCEQgInRuLOrf8Fw0Vsr0MPLm9msuskaL-q1Uw3VK0rDCN1x5t_LOOODhJAqvqqCjxDK2UvOrH7FjktuRScY2zLVgq5tLh-QDZfCdJxr-ZBsWq7veibEBbms9VsLtezNY3LBJVdKKrEhPz8g1LXgjGmheaJ5f7jHRNfdAt-RQvLUQRlzoj7kffBIMYY5JFhCy601pHu6fEU6Huawx_KafoEY_Ok2JAp0xgXGHIOjcW1sDfMaYcnlCXk0Qaz49Hxekc9vbz5dv-_uPr67vX5z1zlh1NKJLZPt_cAsKmGcd6MB22vde4bcyslZKw3j3DvL0E1m6wG1soJLO_VmhP6KvDj13ZX8Y8W6DHOoDmOEhHmtg25dlGW6geYEupJrLTgNuxJmKIeBs-EoffgjffgrffgtvZU-O89Yxxn9v8Kz5QY8PwNQHcSpQHKh_sfp459U_wtAiosb</recordid><startdate>20040601</startdate><enddate>20040601</enddate><creator>ROSENBAUM, Abraham</creator><creator>KIRBY, Christopher</creator><creator>BREEN, Peter H</creator><general>Lippincott</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20040601</creationdate><title>Measurement of oxygen uptake and carbon dioxide elimination using the bymixer: Validation in a metabolic lung simulator</title><author>ROSENBAUM, Abraham ; KIRBY, Christopher ; BREEN, Peter H</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c286t-2405022a09e628cdcb8a93773d0e195fc9958011dc90ecf84dae7692159f38ba3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2004</creationdate><topic>Anesthesia</topic><topic>Anesthesia, Closed-Circuit - instrumentation</topic><topic>Anesthesia, Closed-Circuit - methods</topic><topic>Anesthesia. Intensive care medicine. Transfusions. Cell therapy and gene therapy</topic><topic>Biological and medical sciences</topic><topic>Breath Tests - instrumentation</topic><topic>Breath Tests - methods</topic><topic>Carbon Dioxide - analysis</topic><topic>Carbon Dioxide - metabolism</topic><topic>Humans</topic><topic>In Vitro Techniques</topic><topic>Lung - metabolism</topic><topic>Medical sciences</topic><topic>Oxygen Consumption - physiology</topic><topic>Pulmonary Gas Exchange</topic><topic>Respiration, Artificial - instrumentation</topic><topic>Respiration, Artificial - methods</topic><topic>Ventilators, Mechanical - standards</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>ROSENBAUM, Abraham</creatorcontrib><creatorcontrib>KIRBY, Christopher</creatorcontrib><creatorcontrib>BREEN, Peter H</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Anesthesiology (Philadelphia)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>ROSENBAUM, Abraham</au><au>KIRBY, Christopher</au><au>BREEN, Peter H</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Measurement of oxygen uptake and carbon dioxide elimination using the bymixer: Validation in a metabolic lung simulator</atitle><jtitle>Anesthesiology (Philadelphia)</jtitle><addtitle>Anesthesiology</addtitle><date>2004-06-01</date><risdate>2004</risdate><volume>100</volume><issue>6</issue><spage>1427</spage><epage>1437</epage><pages>1427-1437</pages><issn>0003-3022</issn><eissn>1528-1175</eissn><coden>ANESAV</coden><abstract>The authors have developed a new clinical bymixer that bypasses a constant fraction of gas flow through a mixing arm. A separate bymixer was interposed in the expiratory and inspiratory limbs of the ventilation circuit to measure mixed gas fractions. By utilizing nitrogen conservation, the clinical bymixer allows the determination of airway carbon dioxide elimination (VCO2) and oxygen uptake (VO2), whenever basic expired flow and gas monitoring measurements are used for the patient. Neither an expiratory exhaust gas collection bag nor expensive, complex equipment are needed. This study tested the accuracy of airway bymixer-flow measurements of VCO2 and VO2 in a new bench apparatus. The authors compared airway bymixer-flow measurements of VCO2 and VO2 over a range of reference values generated by ethanol combustion in a new metabolic lung simulator, which was ventilated by a volume-cycled respirator. An airway humidity and temperature sensor permitted standard temperature and pressure, dry, correction of airway VCO2 and VO2. Bymixer-flow airway measurements of VCO2 and VO2 correlated closely (R2 = 0.999 and 0.998, respectively) with the stoichiometric values generated by ethanol combustion. Limits of agreement for VCO2 and VO2 were 0.1 +/- 4.7 and 1.1 +/- 5.7%, respectively. The average (+/- SD) percent error for airway VCO2 (compared with the stoichiometric value) was 0.1 +/- 2.4%. The same error for airway VO2 was 1.1 +/- 2.9%. The new clinical bymixer, plus basic expired flow and gas fraction measurements, generated clinically accurate determinations of VCO2 and VO2. These measurements are helpful in the assessment of metabolic gas exchange in the critical care unit. In contrast to using the gas collection bag or complex metabolic monitor, the bymixer should measure mixed gas concentrations in the inspired or expired limb of the common anesthesia circle ventilation circuit.</abstract><cop>Hagerstown, MD</cop><pub>Lippincott</pub><pmid>15166562</pmid><doi>10.1097/00000542-200406000-00015</doi><tpages>11</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0003-3022
ispartof Anesthesiology (Philadelphia), 2004-06, Vol.100 (6), p.1427-1437
issn 0003-3022
1528-1175
language eng
recordid cdi_crossref_primary_10_1097_00000542_200406000_00015
source HEAL-Link subscriptions: Lippincott Williams & Wilkins
subjects Anesthesia
Anesthesia, Closed-Circuit - instrumentation
Anesthesia, Closed-Circuit - methods
Anesthesia. Intensive care medicine. Transfusions. Cell therapy and gene therapy
Biological and medical sciences
Breath Tests - instrumentation
Breath Tests - methods
Carbon Dioxide - analysis
Carbon Dioxide - metabolism
Humans
In Vitro Techniques
Lung - metabolism
Medical sciences
Oxygen Consumption - physiology
Pulmonary Gas Exchange
Respiration, Artificial - instrumentation
Respiration, Artificial - methods
Ventilators, Mechanical - standards
title Measurement of oxygen uptake and carbon dioxide elimination using the bymixer: Validation in a metabolic lung simulator
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-28T05%3A39%3A06IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Measurement%20of%20oxygen%20uptake%20and%20carbon%20dioxide%20elimination%20using%20the%20bymixer:%20Validation%20in%20a%20metabolic%20lung%20simulator&rft.jtitle=Anesthesiology%20(Philadelphia)&rft.au=ROSENBAUM,%20Abraham&rft.date=2004-06-01&rft.volume=100&rft.issue=6&rft.spage=1427&rft.epage=1437&rft.pages=1427-1437&rft.issn=0003-3022&rft.eissn=1528-1175&rft.coden=ANESAV&rft_id=info:doi/10.1097/00000542-200406000-00015&rft_dat=%3Cproquest_cross%3E71956907%3C/proquest_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c286t-2405022a09e628cdcb8a93773d0e195fc9958011dc90ecf84dae7692159f38ba3%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=71956907&rft_id=info:pmid/15166562&rfr_iscdi=true