Loading…
The non-proportional response of single crystalline oxide scintillators
The energy dependence of the response of single crystalline oxide scintillators (BGO, CWO and GSO(Ce)) was studied over the wide energy range from 6 to 1332 keV. The non-proportionality was defined as the difference between the specific response value at the given energy and the one at 661.6 keV. It...
Saved in:
Published in: | IEEE transactions on nuclear science 1996-06, Vol.43 (3), p.1282-1283 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The energy dependence of the response of single crystalline oxide scintillators (BGO, CWO and GSO(Ce)) was studied over the wide energy range from 6 to 1332 keV. The non-proportionality was defined as the difference between the specific response value at the given energy and the one at 661.6 keV. It is found that a considerable (up to 35%) non-proportionality of the response exists for the all investigated scintillators in the low energy region from 6 to 100 keV, while as in the energy range from 100 to 1332 keV the scintillator's response is proportional to the gamma-rays energy with the accuracy equal to /spl plusmn/1%. |
---|---|
ISSN: | 0018-9499 1558-1578 |
DOI: | 10.1109/23.506678 |